Safety Instructions - IKA Vibrax VXR basic Mode D'emploi

Masquer les pouces Voir aussi pour Vibrax VXR basic:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Contens
Declaration of conformity
Waranty

Safety instructions

Proper use
Unpacking
Useful facts
Commissioning
Motor protection
Maintenance and Cleaning
Accessories
Technical data
Safety instructions
Read the operating instructions in full before starting up and fol-
low the safety instructions.
Keep the operating instructions in a place where they can be accessed by
everyone.
Ensure that only trained staff work with the appliance.
When adjusting the speed, pay careful attention to the containers fastened onto
the agitation table to prevent the medium to be shaken from possibly splashing
out. If the target value of the speed is changed too rapidly, the machine will turn
itself off automatically.
Please observe the relevant safety information and guidlines as well as work
protection and accident prevention requirements for use in the laboratory.
Avoid allowing objects to push or strike the agitation table. Even small unde-
tectable damage may result in serious damage to the motor bearing. Careful
handling will guarantee safe work and a long service life of the machine.
If you notice that the device is not running smoothly, the speed must always be
reduced until no more unevenness occurs in the operation. Please make certain
that individual agitation containers are placed in the middle of the agitation table
and that if several containers are placed on the agitation table together, they are
evenly distributed and well secured.
EN
Because of improper loading and the position of the center of gravity, dynamic
Page
forces may arise during the agitation process that cause the shaker to move
2
about on the table. In this case, reduce the speed immediately.
2
Additional risk may be present for the user if flammable materials are used during
7
the agitation process or if mechanical agitation energy is transferred to glass
8
equipment that is being used, causing it to break.
8
8
The holder and attachment used for the agitation process must be fastened
8
securely on the agitation table. Tighten all three clamping screws are firmly in
9
place from time to time, since they could come loose during operation. Also
9
make certain that the agitation vessels used during the agitation process are
9
securely in place on their respective agitation attachment.
10
Before placing the device in service, please ensure that the turn dial for adjusting
the speed is against the left stop, since the device will begin to run at the last
speed that was set.
After an interruption in the power supply during an agitation procedure, the
device will start running again by itself.
Only a specialist may open the device.
Covers or parts that can be removed from the device must be put
in place on the device in order tro ensure safe operation.
The IKA Vibrax VXR was not constructed for operation in dangerous atmosphe-
res, for mixing dangerous substances or for operation under water.
The instrument may not be operated in rooms with explosion hazards.
Furthermore, it has to be placed upon a fire-proof and/or not burnable surface.
Use only original IKA accessories.
7

Publicité

Chapitres

Table des Matières
loading

Table des Matières