Eaton MGR10 Serie Manuel D'utilisation page 71

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

DATAlogger:STEP
Forces a single measurement only, which is stored in the next datalog location. The
microhmmeter then returns to single trigger mode.
If the log has reached the COUNt value, any further STEP commands will generate an
'execution error'. The command will be ignored, and the microhmmeter will remain in the
single trigger mode.
DATAlogger:VALue? <value number|ALL>
Transfers a datalog reading from the memory to the output buffer. A value number of
between 1 and POINts will output one reading from that location. If ALL is used, all
readings are sent sequentially to the output buffer from location 1 to POINts. Sending
ABORt will cancel the operation. The output format is:
The "range" will include the letter 'z' if the Zero function was used on a measurement, and
the letter 'T' if Temperature Compensation was used.
DATAlogger:POINts?
Returns the number of readings actually stored in the datalog memory. The response will be
between 0 and COUNt.
CALCulate:DATA Group
These commands perform statistical calculations on the current contents of the data log. A
minimum of 2 readings must be stored otherwise an 'execution error' is generated, i.e. bit (4) of
the Standard Event Register is set. This error is also generated if the log contains inconsistent
entries, e.g. readings taken on different ranges.
CALCulate:DATA:MINimum?
Returns the minimum resistance value stored in the data log.
CALCulate: DATA:MAXimum?
Returns the maximum resistance value stored in the data log.
CALCulate: DATA:AVERage?
Returns the average of all resistance values stored in the data log.
CALCulate: DATA:PTPeak?
Returns the peak-to-peak value of the resistance values stored in the data log, i.e. max to
min.
CALCulate: DATA:SDEViation?
Returns the standard deviation of the resistance values stored in the data log. This is
<record no.>,<"range">,<resistance>,<"date">,<"time">