Bluephase Meter-BA-WE3_611504_Rev4_bluephase meter.qxd 24.10.12 08:12 Seite 35
Appendix
Guidance and manufacturer's declaration - electromagnetic emission
The "Bluephase Meter" is intended for use in the electromagnetic environment specified below. The customer or the user of the "Bluephase Meter"
should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3

Guidance and manufacturer's declaration - electromagnetic immunity

The "Bluephase Meter" is intended for use in the electromagnetic environment specified below. The customer or the user of the "Bluephase Meter"
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV line - line
IEC 61000-4-5
± 2 kV line - earth
Voltage dips, short
<5 % U
T
interruptions and
(>95 % dip in U
voltage variations on
40 % U
T
power supply input
(60 % dip in U
lines
IEC 61000-4-11
70 % U
T
(30 % dip in U
<5 % U
T
(>95 % dip in U
Power frequency
3 A/m
(50/60 Hz) magnetic
field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
No power network input
No power network input
Compliance level
± 6 kV contact
± 8 kV air
n.a. (no cables connected)
n.a. (no cables connected)
) for 0.5 cycle
T
) for 5 cycles
T
n.a. (no mains power)
) for 25 cycles
T
) for 5 sec
T
30 A/m
Electromagnetic environment - guidance
The "Bluephase Meter" uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
The "Bluephase Meter" is suitable for use in all establishments, including domestic establish-
ments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Electromagnetic environment - guidance
Floors should be concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Power frequency magnetic fields should be at levels characteristic of a typical location
in a typical commercial or dental environment.