Rinse the medicine cup with distilled water after cleaning with citric acid
solution. To do this, pour the distilled water into the medicine cup to a
level of 6 ml, close the lid and shake the cup at least 5 times. Then open
the lid and pour out the water. Rinse the outer part of the membrane
with distilled water to remove citric acid residues. Repeat the cleaning
procedure with distilled water at least 3 times.
2. Disassemble the nebulizer.
Remove the inhalation mask or mouthpiece from the nebulizer (2.1).
Remove the cover (2.2).
Remove the medication chamber from the nebulizer (2.3).
2.2
WARNING
Before the first use, the inhalation mask should be rinsed with water and dried.
Some of the components of the device can be re-used repeatedly after
disinfection (see how to disinfect on page 119).
3. Rinse the parts with enough
distilled water.
Rinse the medication chamber, mouthpiece, inhala-
tion mask and liquid condenser with water.
EN
2.1
Magnetic
lock
2.3
Press to release
medication chamber
119