AGU Baby Tomchi AGU N8 Mode D'emploi page 118

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3. Turn off the power.
The nebulizer automatically turns off
after 15 seconds after out of drugs.
If you want to stop the inhalation, press
the «I/O» button to turn off the power.
Press the «I/O» button.
DEVICE CLEANING AFTER USE
Be sure to clean the medicine cup with distilled water after each use before
transportation or storage.
WARNING
1. Remove any medicine residues.
Open the lid of the medicine cup and remove any medicine residues.
Use citric acid solution to clean the medicine cup*. Method of preparation
of citric acid solution: pour 100 ml of drinking water into a clean container
or glass. Add half a teaspoon of citric acid (approx. 2.5 - 3.0 g) to 100 ml
of drinking water and mix thoroughly. Pour the prepared citric acid solu-
tion into the medicine cup to a level of 4 ml and close the medicine cup lid.
*Use crystalline or powdered citric acid to prepare the solution.
Press and hold the ON/OFF button for 5 seconds. The device will turn on
and switch to cleaning mode, and the LED indicator of the medicine cup
will start flashing blue at a rate of 3 times per second, the network indica-
tor will start flashing orange against a background of constant green light.
Spray a citric acid solution for two minutes to remove the medicine resi-
dues. If necessary, repeat the above procedure until the medicine resi-
dues are completely removed.
If the citric acid solution in the medicine cup runs out and the device
produces a high-frequency sound, press the ON/OFF button to turn off
the device. Otherwise, the membrane may wear out.
Please remove the medicine residues after each use. Otherwise, the
medicine cup membrane may become clogged.