To do so the following measures are required:
Always operate the Analyzer with the AC adapter while it is
not used in portable mode. A permanent connection to the
mains power is not detrimental to the integrated battery but
maintains its performance.
Connect unit to the 12 V supply cable when using in
vehicles.
For portable operation with low battery indication, connect
the charger and fully charge battery promptly.
Keep the unit connected to the charger when storing.
For long term storage of the unit without the possibility to
connect the battery charger, charge the Analyzer
completely before storage and fully recharge the battery
every 8 weeks.
6.4

Maintenance work

We recommend that the unit be serviced on a regular basis
every 12 months.
Contact Dräger or a service organization authorised by Dräger
for maintenance.
We recommend concluding a service agreement.
7
Disposal
Dispose of the product in accordance with the applicable rules
and regulations.
Disposing of electric and electronic equipment
According to Directive 2002/96/EC, this product
must not be disposed of as municipal waste. It is
therefore labelled with the symbol depicted here. The
product can be returned to Dräger free of charge.
Please contact your national Dräger Sales
Organisation or Dräger for more information.
Dräger DrugTest 5000 Analyzer
8

Technical data

Supplyvoltage
12 VDC (11 to 15 VDC)
Power input
approx. 3 A
Dimensions
200 mm x 250 mm x 220 mm
(W x H x D)
Weight
4.5 kg (including battery)
Temperature
Operation
5 °C to 40 °C (+41 °F to +104 °F)
Storage and
–20 °C to 60 °C (–4 °F to 140 °F)
transport
Ports
optical IRDA (printer),
PS/2 (for external keyboard or
scanner),
USB Slave (for communication with
a PC)
Measurement duration
four minutes and 15 seconds
Storage capacity
500 data sets with test results
9

Order list

Name and description
Dräger DrugTest 5000 Analyzer
including battery charger supply unit, mains
cable and Instructions for Use
Keyboard, compact
(PS/2,
dimensions approx. 28.2 x 13.2 x 2.4 cm):
"QWERTY" – key assignment English
"AZERTY" – key assignment French
Dräger Mobile Printer
Paper for Dräger Mobile Printer
(5 rolls)
Carrier bag
Dräger DrugTest 5000 Analyzer
Charger power supply cable (12 V)
with mains cable for EU, UK, USA, AUS
Vehicle supply cable (12 V)
Power cable from car on-board/cigarette
lighter socket to Dräger DrugTest 5000
Analyzer
USB connection cable
for communication with a PC
Dräger DrugTest 5000 test kits, e.g.:
Dräger DrugTest 5000 STK-CA
Package with 20 pieces
Test kit for instructing users in the use of the
Dräger DrugTest 5000 test system, also for
quality analysis.
Package with 6 pieces (3 x Positive; 3 x
Negative)
Dräger DrugTest 5000 System Technical
Handbook
Dräger Diagnostics
Software for Analyzer configuration
Disposal
Part number
3701413
with
CFG 83 26 649
83 15 497
83 15 142
83 19 310
83 19 002
83 25 333
83 15 675
83 12 166
AG 02 661
3700791
83 23 640
91 00 163
contact Dräger
Services