Use
4.1
Prerequisites
NOTICE
i
i
Observe the Instructions for Use of the utilised Dräger
DrugTest 5000 test kits.
The Dräger DrugTest 5000 Analyzer is delivered as a fully
operational unit. Initial operation does not require any prior
assembly or installation.
4.2
Preparation for use
Place the Dräger DrugTest 5000 Analyzer on a flat, firm
and horizontal surface.
Keep all air vents uncovered and clean.
Ensure the specified ambient conditions:
Ambient temperature between 5 °C and 40 °C
(41 °F and 104 °F)
Relative humidity between 5 % and 95 % r.h.
When the unit is carried from a cold environment to a
warmer place, condensation may occur.
Wait for the Analyzer to adjust to the environmental
temperature and make sure it is dry before activating it.
Wait
approximately
temperature changes from –20 °C to +20 °C (–4 °F and
+68 °F).
Use any of the available accessories such as the external
keyboard or set up the printer.
Connect the battery charger if used with an external power
supply, e.g. if the battery charge is insufficient.
4.3
Switch on Analyzer
Keep the
key pressed until the start window appears.
OK
After successful completion of the automatic self check, the
unit displays » Ready « to indicate its readiness for
operation.
4.4
Collecting the sample
NOTICE
i
i
Observe the Instructions for Use of the utilised Dräger
DrugTest 5000 test kits.
4.5
Analysing the sample
1. Make sure that Analyzer is switched on and » Ready «.
2. Take the test cassette with the sample from the test
subject.
Open the Analyzer door and insert the test cassette into the
bottom slot of the Analyzer until it audibly clicks into
position (Fig 4).
3. Remove safety cap from the cartridge and insert cartridge
into the top slot of the Analyzer until it audibly clicks into
position (Fig 5).
4. Close the door.
The Analyzer now automatically starts the analysis.
To show results follow the instructions on the screen.
On completion of the analysis, the Analyzer displays the
results for every tested substance on the screen.
Note the display!
Dräger DrugTest 5000 Analyzer
1.5
hours
if
the
ambient
NOTICE
i
i
Do not move the unit during the test evaluation!
A status bar on the display indicates the progress of the
evaluation process. Once the evaluation process is
complete, the results are displayed for each tested
substance.
Optional:
4.5.1
Entering data
Data of the subject and user can be entered, the test result
can be printed and the data can be saved by selecting the
corresponding control panel keys. Follow the instructions
on the display.
4.6
After the analysis
The unit emits an acoustic signal at the end of the analysis.
1. Remove test cassette with attached cartridge from the
Analyzer after the analysis and dispose of it (Fig 6).
2. Close the door of the Analyzer.
The analysis results are depicted on the display.
Optional:
Printing the current results.
3. Press the
key to confirm results.
OK
The Analyzer is now automatically prepared for the next
measurement.
4.7
Switching Analyzer off
1. Keep the
OK
key pressed. The unit will switch off after
approx. 3 seconds. The power supply cable should remain
connected to charge the battery until the unit is re-packed
for transport.
NOTICE
i
i
To maintain the performance of the battery, please
read the battery care notes, see "Battery maintenance"
on page 6.
In order to protect the optical components of the
Analyzer against dust, keep the door of the Analyzer
closed at all times unless you are inserting or removing
test kits.
4.8
Operating modes
The Dräger DrugTest 5000 Analyzer has three operating
modes:
Ready
The unit is ready to measure after switching on and running the
successful self test. The three keys and the display of the unit
light up.
Standby
After a predefined period of inactivity in ready mode, the unit
switches to standby. The keys on the unit are still lit; however,
the display is switched off. You can quit standby mode and
return the unit to immediate ready mode by pressing any key.