The CORSIA atomiser does not need maintenance or lubrication.
Before use always check correct functioning and safety of the device. Carry out disinfection as described in the "CLEANING
ACCESSORIES" section. Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have
been damaged during previous use.
Connect cable to electrical network and turn switch on.
Close the air outlet with one finger to make sure that noise produced is regular and there is no malfunctioning.
With the air outlet always closed check the correct functioning of the nebulization regulator by turning the knob form MIN to
MAX. Make sure the indicator of the pressure meter is working correctly.
Verify that the atomiser is not damaged by previous use (it was badly put away or badly knocked).
A protection fuse (F 1x1.6A 250V) reachable from exterior and it situated in the plug protects the instrument.
For use replacing, always check the type and the range indicated.
Fault type
1. The device doesn't work
2. Low Nebulization
3. Low Nebulization
4. Absence of Nebulization
5. Slow Nebulization
6. Noisy Device
Fault 1 - 2 - 3 - 4 - 5 - 6
the unit does not nebulizer once the above conditions have been checked, we suggest to contact your dealer or technical
If
service GIMA S.p.A.
The manufacturer will provide upon request electric diagrams, component list, descriptions, setting instructions and
any other information that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
GIMA S.p.A. DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICL SERVICE CHECKING, APPEARS
MAINTENANCE
Cause
a) The plug may be misplaced in the wall
socket
b) Thermal protector may be on (the device has
been working beyond its limits and / or near
heat sources)
Clogged Nebulizer Tank
Clogged Nebulizer Tank
Clogged Nebulizer Tank
Air tube is bended or squeezed
Highly dense drug
Extended use
No solution with previous items
GIMA S.p.A. TECHNICAL SERVICE
TO BE TAMPERED
a)
Make sure the plug is properly placed in the wall
socket. Make sure the ON/OFF switch is in position I.
b)
Switch off the device by pressing the switch to
position 0 and left the motor cool down for at least 30
minutes.
Clean and disinfect the nebulizer tank as explained in the
instruction manual
If cleaning was not succesful change cruet
a)
Check that the nebulizer contains medication; Make
sure that the nebulizer is not clogged;
Check the connection between the compressor air
outlet port and the accessories
b)
Make sure the air tube is not bended or squeezed.
Dilute drug in physiological liquid
Call retainer or manufacturer GIMA S.p.A.
Call retainer or manufacturer GIMA S.p.A.
1-
Air Tube
2-
Nebulizer Tank
3-
Nebulization Nozzle
4-
Nebulizer Top
5-
Mouthpiece
6-
Adult Mask
7-
Pediatric Mask
8-
Nosepiece (accessory under
request)