Actuating the cycle switch to the TIMED position initiates the stomaching process,
which will continue until either: -
7B.
Setting the cycle switch to the CONTINUOUS position initiates the stomaching process,
which will continue until either: -
WARNING:- Opening the door during the operating cycle causes the instrument to
8.
After processing, simply open the door and remove the Stomacher Bag(s). The
instrument is immediately ready for processing the next sample.
NOTE: -
A motor electrical current limiting device is incorporated into the control circuit. Upon initial start
up a short time may elapse before the motor builds up torque and runs normally. This condition is
more likely to prevail under the following circumstances: -
i)
ii)
For condition (i) run the instrument for 5-10 minutes, preferably with a number of stop-start
cycles. For condition (ii) allow the instrument to stand in a warm environment as detailed in
Section 8.3, followed by a short period of running with stop-start cycles.
Issue: 3.2
STOMACHER
USER INSTRUCTIONS
The timed cycle has timed out.
The CYCLE switch has been set to the STOP position.
The door has been opened.
The power has been switched off.
The CYCLE switch has been set to the STOP position.
The door has been opened.
The power has been switched off.
stop. However, this must not be considered as the normal method for
stopping the instrument.
Upon initial start up following a period of transportation or storage.
The instrument has been stored or located in a cold environment (less than 10°C
allowing the rubber connectors to become stiff).
®