STOMACHER
7.2.3
Manual Operation
Please note that pressing any control function key is acknowledged by a short tone (0.1 seconds
duration) emitted from the instruments internal sounder.
1.
Ensure the POWER rocker switch is in the on (1) position. The green lamp in the
switch should be illuminated indicating that power is present and the instrument
should also emit a short tone. The display should now indicate the instrument
status.
2.
Ensure the AUTO icon is not displayed. If AUTO is shown, press the AUTO key
to revert to MANUAL mode. Similarly, if programmes, P1, P2, or P3 are selected,
press PROG repeatedly until no programme is selected.
3.
Open the door by raising the lid fully upward and back. This gives access to the
paddle compartment. The DOOR OPEN icon will now be displayed.
4.
Place the Stomacher
protruding above the bag clamp and close the door thus clamping the bag(s). The
DOOR OPEN icon will now be off.
5.
The ENUNCIATOR icon indicates that the time parameter has been selected
(display defaults to 30 seconds). The time parameter can now be adjusted by
pressing the + or – key as required. The setting is variable between 0 seconds and
99 minutes 59 seconds in increments of 1 second. Pressing and releasing the + or –
key increments the time by 1 second, holding down the + or – key accelerates the
rate of change.
6.
Pressing the SPEED / TIME key toggles the ENUNCIATOR icon to indicate that
the speed parameter has been selected (display defaults to 230rpm). The speed
setting can now be adjusted by pressing the + or – key as required.
7.
Pressing the START key will initiate the stomaching process. The instrument will
run at the selected speed and duration. Please note that the time display counts
down from the selected time and displays the remaining processing time. The
stomaching process continues in normal circumstances until either: -
In all instances the time resets to the selected value.
8.
After processing, simply open the door and remove the Stomacher
instrument is ready for processing the next sample.
WARNING: - Opening the door during the operating cycle causes the instrument
7.2.4
Automatic Operation
Issue: 3.2
®
3500 AND 3500 BIOWASHER
USER INSTRUCTIONS
®
Bag(s) into the paddle compartment leaving 50 to 60mm
The timed cycle has finished.
The STOP key is pressed.
The door has been opened.
The power has been switched off.
to stop. This, however, must not be considered the normal method
for stopping the instrument.
®
Bag(s). The