Pièce à main Dépannage
Pièce à main haute vitesse
PROBLÈME
Cause
SOLUTION
Pas de débit d'eau dans la pièce à main
1 Compresseur pas sur ON.
2. contrôle du débit ne ON (P4-E).
3 Eau / CLS bouteille vide.
4 Vérifier un «coude» dans la ligne d'eau.
5 Vérifiez si 3 voies seringue a l'eau.
6 L'utilisation de l'eau du robinet, le blocage des vannes d'eau.
7. basse pression du régulateur.
1. Turn compressor ON and check that ON/ OFF tap is in the ON position.
2. Turn water control to ON.
3. Fill water/CLS Bottle(s).
4. Check lines and pressure.
5. If yes, lack of water must be in the valves or the handpiece. The 3 way syringe is the quickest way to check for water.
6. Valves will need replacing.
7. Check regulator is set at 70-80 PSI.
Mineral Accumulation
1. Use of tap water - mineral or chemical action blockage will usually take 3-4 years to occur.
2. Verdigrease (Green Slime) build up in brass water valves due to chemical reaction with tap water.
1. Pass a fine wire through the water intake hole. Pass another fine wire through the hole in the handpiece head. Pass the
wire as far as possible each way. 'Blow out' the loosened mineral deposits.
2. Replace valves.
Use of Chlorhexidine
1. Most brands of Chlorhex will de-posit out of solution and block valves, etc. CLS is designed to be used in its diluted
strength without depositing.
NOTE: Do not dilute CLS beyond its label recommendations or it could result in deposits.
2. Chlorine in tap water will cause a flocculation and block filters and valves
1. Contact your local iM3 Official Service Center.
2. Contact your local iM3 Official Service Center. Use only distilled water
Cartridge Toggle Toggle switches become floppy All iM3 dental units use a unique cartridge system for the toggle switches. The toggles are used to pressurize the water/CLS bottles and
select the low or high speed handpiece. When the o-rings be-come worn you simply push the cartridge in and turn a 1/4 turn anti clockwise. The cartridge can then be removed from the
valve casing. To install a new toggle, locate the desired position, push and rotate a 1/4 turn clockwise. New toggles can be ordered from iM3 and installed by the user. Toggles can be
installed in the 12, 3, 6 and 9 o'clock positions. If the toggle is in the wrong position simply remove and reposition. View this procedure on the iM3 website