TROUBLESHOOTING
Main Display reads:
"ovEr" or "undr"
Secondary Display reads:
"undr"
"ovEr"
SPECIFICATIONS
Display
Dissolved O
2
(ppm or mg/L)
Dissolved O
2
% air-sat
Temperature °C
Pressure Compensation
Salinity Compensation
ATC Probe
Calibration Backup
Audio Feedback
Power Source
Operating Temperature
Instrument Casing
Weight (with battery)
Dimensions (W x D x H)
CONVERSIONS
To Convert:
Inches of Hg to mBar
Inches of Hg to mmHg
mmHg to mBar
RECOMMENDED SPARE PARTS LIST
PART #
200-4
4 meter probe and cable assembly
200-10
10 meter probe and cable assembly
280
DO carrying case, hard sided
5908
Membrane kit, 1.25 mil PE (605306), six cap membranes and KCl solution
405
Instrument carrying case, soft
• Check membrane and electrolyte solution.
• Clean anode and cathode.
• Return product for service.
• Heat the sample to above –6.0 °C
• Return product for service.
• Cool sample to below 46.0 °C
• Return product for service.
Range
0 to 20.00 ppm (mg/L)
0 to 200.0 %
-6.0 to 46.0 °C
600 to 1100 mBar (450 to 825 mmHg)
From 0.0 to 40.0 ppt
Thermistor, 10KΩ, at 25°C
Yes
Yes, on all keys
One 9V battery
0 to 50°C
Splash-resistant IP 65
350 grams (.75 lbs)
186 mm x 70 mm x 37 mm (7.3 in x 2.8 in x 1.5 in)
Multiply by:
33.864
25.4
1.333
Possible Solutions:
Possible Solutions:
Accuracy
±2 % of the reading or ±2% air
saturation, whichever is greater
±2% of the reading or ±0.2 ppm,
whichever is greater.
±0.3 °C ±1 digit
DESCRIPTION
Item #605368 • Drawing #A605368
6
Resolution
0.01 mg/L
0.1 %
0.1 °C
Revision A • December 2002