### Sample Information ###
Sample Type:
PEG+PPG+Raffinose
### Tuning Condition ###
Detector Adjustment:
On
Resolution Adjustment:
On
FWHM of Spectrum:
0.70
Sensitivity Adjustment:
On
Mass Calibration:
On
### Tuning Result ###
Model:
LCMS-8045
Interface:
DUIS
Polarity:
Neg
Tuning Mode:
Auto
Tuning Date:
5/17/2017 11:13:40 AM
Acquisition Mode:
Product Scan
Nebulizing Gas Flow:
1.50(settings : 1.50) L/min
Drying Gas Flow:
0.00(settings : 10.00) L/min
### Tuning Profile ###
CD:+10.00 kV
CD:+10.00 kV
Int.: 578546
Int.: 727733
FWHM: 0.73
FWHM: 0.75
m/z : 89.00
m/z : 179.05
Pre m/z: 503.15
Pre m/z: 503.15
Mag.: 34
Mag.: 27
Scan Range: 2.00 - 553.15 Scan Speed: 30 u/s Base Peak: 221.05 (793507) Pre m/z: 503.15
100
179.10
89.00
100
200
m/z(Target)
m/z(Actual)
Difference
503.15 > 89.00
89.00
503.15 > 179.05
179.10
Const
65.05
DL Bias
Const
+0.0
+0.0
Qarray Bias
Const
+0.0
+0.0
Qarray RF
Scan
+50.0
+19.3
Multipole RF
Scan +100.0
+30.0
Multipole1 Bias
Const
+0.0
+0.0
Multipole1 Lens
Const
+0.8
+0.8
Multipole2 Bias
Const
+0.8
+0.8
Multipole2 Lens
Const
+1.0
+1.0
Q1 Bias
Const
+3.5
+3.5
CC Lens1
Const
+0.0
+0.0
CC Lens2
Const +100.0
+100.0
CC Lens3
Const
+20.0
+20.0
Const
45.05
CC RF
Scan +100.0
+100.0
CE
Scan
+30.0
+30.0
Q3 Bias
Scan
+7.0
+7.0
Const 107.15 1071.50 1893.40
Q1 Pre Bias
Scan
+15.0
+5.0
Q3 Pre Bias
Scan
+15.0
+5.0
CID Gas Flow:
Heating Gas Flow:
Interface Bias:
Interface Current:
Interface Temp.:
DL Temp.:
Heat Block Temp.:
Q1 RF Gain:
Q1 RF Offset:
Q3 RF Gain:
Q3 RF Offset:
Q1 Post-rod Bias:
CID CELL Exit Lens:
Conversion Dynode:
Detector:
PG Vacuum:
IG Vacuum:
Corona Needle Voltage:
Corona Needle Current:
300
400
500
Width
Intensity
0.00
0.73
578546
0.05
0.75
727733
168.10
256.15
344.20
652.40 1004.60 1893.40
+0.0
+0.0
+0.0
+0.0
+0.0
+0.0
+0.0
+0.0
+50.0
+90.0
+90.0
+130.0
+50.0
+50.0
+50.0
+300.0
+0.0
+0.0
+0.0
+0.0
+0.8
+0.8
+0.8
+0.8
+0.8
+0.8
+0.8
+0.8
+1.0
+1.0
+1.0
+1.0
+3.5
+3.5
+3.5
+3.5
+0.0
+0.0
+0.0
+0.0
+100.0
+100.0
+100.0
+100.0
+20.0
+20.0
+20.0
+20.0
89.05
133.10
168.10 1893.40
+120.0
+300.0
+300.0
+300.0
+30.0
+30.0
+30.0
+30.0
+7.0
+7.0
+7.0
+9.0
+50.0
+50.0
+50.0
+50.0
229(settings : 230) kPa
4.97(settings : 5.00) L/Min
-3.00 kV
0.71 uA
100(settings : 100) C
250(settings : 250) C
399(settings : 400) C
5000
5000
4999
5004
5.0 V
-0.7 V
+10.00 kV
-2.28(-2.28) kV
7.2e+001 Pa
2.1e-003 Pa
-3.50 kV
0.10 uA
600
700
800
+0.0
+0.0
+0.0
+0.0
+130.0
+130.0
+300.0
+300.0
+0.0
+0.0
+0.8
+0.8
+0.8
+0.8
+1.0
+1.0
+3.5
+3.5
+0.0
+0.0
+100.0
+100.0
+20.0
+20.0
LCMS-8045
3.8 Réglage automatique
3
900
99
LCMS-8050
CL
CL