______________________________________________________________
3.
SPECIFICATIONS
Measuring ranges
Output sensitivity
Accuracy (45-65Hz)
Minimum load
Frequency range (-2dB) 20Hz to 100kHz (ACP 2005)
Phase (45-65Hz)
Position sensitivity
External field
Noise (ACP 2005)
(ACP 2015)
Temperature coeff.
Power supply
Battery life
External Power
Low battery
Overload
Enclosure
Material
Degree of Protection
Dimensions
Output
_______________________________________________________________
5
20A / 200A / 2000A AC
100mV / 10mV / 1mV per A (AC coupled)
± 1% of reading ± 0.2A 20/200A ranges
± 1% of reading ± 1A
100kΩ for specified accuracy
20Hz to 1MHz (ACP 2015)
< ± 1°
± 2.5 % of reading
± 1% of range with cable >100mm (4")
from the probe
0.03A / 0.075A / 0.075A (20A/200A/2000A range)
0.08A / 0.6A / 6A (20A/200A/2000A range)
± 0.15% of reading/° C
Two AA MN1500 LR6 alkaline batteries
or an external power supply (+3V/100mA)
150 hours typical ACP 2005
100 hours typical ACP 2015
Via 2.1mm (0.08") mini power connector
Indicated by an orange LED
Indicated by a red LED
ARNITE, UL94 V0
IP40 (not certified by UL)
110 (l) x 65 (w) x 23 (d) mm
0.5m coax cable terminated with a safety BNC or
0.5m coax cable terminated with a safety 4mm plugs
2000A range