Ulthera Ultherapy Mode D'emploi page 57

Masquer les pouces Voir aussi pour Ultherapy:
Table des Matières

Publicité

N
O
P
S
T
U
Ulthera, Inc.
Code N
USB flash memory connectivity
Please check flash drive and
continue.
See User's Manual for further
information.
Code O
The system scanning is disabled
because of multiple Code Ms.
Please contact Ulthera Support
Code P
Hardware halted
Please restart the system.
If the problem persists please
see the User's Manual for further
information or contact Ulthera
Support.
Code S
The red STOP button has been
pressed.
Please restart the system.
If the problem persists please
see the User's Manual for further
information or contact Ulthera
Support.
Code T
Internal handpiece temperature
is too high
If the problem persists please
see the User's Manual for further
information or contact Ulthera
Support.
Code U
Control unit temperature too
high
If the problem persists please
see the User's Manual for further
information or contact Ulthera
Support.
Un problème a été détecté avec la clé
USB connectée à l'appareil. S'assurer
que la clé a été formatée
correctement et qu'elle dispose de
suffisamment de mémoire. Ne pas
retirer la clé USB pendant que le
système communique avec elle.
Le système a connu de nombreux
codes M et a été verrouillé. Ceci
permet une évaluation plus poussée
par Ulthera afin de minimiser le risque
de dommages.
L'équipement a cessé de fonctionner
en raison d'un événement détecté
dans l'unité de commande.
Le bouton rouge d'arrêt d'urgence a
été enfoncé.
La température interne du dispositif à
main dépasse la limite prévue. Laisser la
tête émettrice refroidir ou en utiliser une
autre.
La température interne de l'unité de
commande dépasse la limite prévue.
Laisser l'unité de commande refroidir.
Fournir une ventilation adéquate.
Page
57

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

System

Table des Matières