ergo vac compact
10

EMC valeurs

The unit meets the Collateral Standards of Electromagnetic compatibility – Requirements and tests
IEC/EN 60601-1-2 the limits and methods of measurement of electromagnetic disturbance characteristics of
industrial, scientific and medical radio frequency equipment.
Medical electrical equipment is subject in regard to the electromagnetic compatibility (EMC) and its special
precautionary measure. The unit must in reference to the mentioned EMC-hints in the accompanying documents
be installed and operated.
This medical device is intended for use in the electromagnetic environment specified in the following tables.
The user of this device should ensure that it is used in such an environment.
Guidance and manufacturer's declaration – electromagnetic emissions
The ergo vac compact is intended for use in the electromagnetic environment specified below.
The customer or the user of the ergo vac compact should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker
emissions
IEC 61000-3-3
page 38
Mode d'emploi
Compliance
The ergo vac compact uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The ergo vac compact is suitable for use in all establishments, including do-
Complies
mestic establishments and those directly connected to the public low-voltage
inherently
power supply network that supplies buildings used for
domestic purposes.
Complies
inherently
Electromagnetic environment - guidance