I N T E L E C T ® M O B I L E 2 Q U I C K S TA RT G U I D E
APPENDIX
ULTRASOUND SPECIFICATIONS
Frequency
Duty Cycles
Pulse Repetition Rate
Pulse duration:

OUTPUT POWER

1 cm2
US applicator
Frequency
1MHz 3MHz 1MHz 3MHz 1MHz 3MHz
Effective
Radiating Area
1
0.9
ERA INTL (cm2)
Max Output
power in
2W
1.8W
Continuous mode
Max Output
2.7W
power in Pulsed
3W
(*)
mode
Max Amplitude
2W/
2W/
in Continuous
cm2
cm2
mode
Max Amplitude in
3W/
3W/
Pulsed mode
cm2
cm2
(*) An error of + 0.25 W can be measured with 1cm2
US applicator, pulse mode 100Hz at 10% or 20% Duty
Cycle.
18
1 MHz; 3 MHz
10%, 20%, 50%, Continuous
16, 48, or 100 Hz
1 -31�25 ms
- Max (ON): 31�25 ms
- Min (OFF): 5ms
2cm2
5cm2
1.5
1
2.5
2.7
3W
2W
5W
5.4W
4.5W
3W
7.5W
8.1W
2W/
2W/
2W/
2W/
cm2
cm2
cm2
cm2
3W/
3W/
3W/
3W/
cm2
cm2
cm2
cm2
Unless otherwise specified, ultrasound
controls accuracy is:
Peak to Average Ratio:
Beam Nonuniformity Ratio
Beam Type
Treatment Time
GENERAL SYSTEM OPERATING AND

STORAGE TEMPERATURE

Operating Conditions
The device will meet its requirement under the
following conditions:
Temperature:
Relative Humidity:
Atmospheric Pressure:
Transport and Storage Conditions
The device will remain in proper condition under the
following conditions:
Temperature:
Relative Humidity:
Atmospheric Pressure:
± 20 %�
1:1, at 50% Duty Cycle
4:1, at 20% Duty Cycle
9:1, at 10% Duty Cycle
<5:1
Collimating
1 to 30 min
5°C to 40°C
15% to 90%
70kPa to 106kPa
-20°C to 60°C
10% to 90%
50kPa to 106kPa

Publicité