MAINTENANCE & CLEANING MANUAL
1600111
SABLE 20K E-Type 4-Hole Mini Air Motor
1600110
SABLE 20K E-Type 4-Hole Air Motor
1610110
SABLE 20K E-Type 4-Hole Air Motor
with Internal Air/Water
1600102
SABLE 5K E-Type 4-Hole Air Motor
1600005
SABLE 5K U-Type Hygienist
Prophy Swivel Handpiece
SPECIFICATIONS
Model
1600111
1600110
SABLE 20K
SABLE 20K
E-Type 4-Hole
E-Type 4-Hole
Mini Air Motor
Air Motor
Handpiece
Slow Speed Air
Slow Speed Air
Type
Driven Motor
Driven Motor
Attachment
ISO 7785-2
ISO 7785-2
Type
Type and
Type and
OEM type
OEM type
Rotation
Up to 20,000
Up to 20,000
Speed
RPM
RPM
Air Pressure 3 bar =43.5 PSI 3 bar =43.5 PSI 3 bar =43.5 PSI 3 bar =43.5 PSI 3 bar =43.5 PSI
Weight
90 grams
90 grams
Sterilization Autoclave
Autoclave
135°C @3min,
134°C @8min,
dry time =16min
dry time=20 min
Sterilization
250 + as per
250 + as per
Cycles
ISO 7785-1
ISO 7785-1
SPECIAL PRECAUTIONS:
1. Always exercise extreme caution when operating your Motor or Prophy
Handpiece.
2. Always inspect your Motor or Prophy Handpiece thoroughly before each use.
3. Always use PPE (Proper Protective Equipment) when operating and cleaning
your Sable Motor or Prophy Handpiece.
4. See your Occupational Health and Safety Act or other safety ordinances for
specific instructions.
5. This is a precision instrument. Do not drop or apply excessive force when
handling.
6. If product is dropped your dealer should be contacted, it is recommended that
you do not use.
7. Contact your dealer right away if any problem should occur with any of your
Sable products.
8. Sterilize before first time use and before every subsequent use thereafter.
9. Indication for use: Used by a trained dental practitioner in conjunction with other
attachments for the cleaning of teeth.
HOW TO USE:
1. Attachment of E-Type Motor or Prophy Handpiece. Align the pins on the Motor
or Prophy Handpiece with the holes in the Air tubing turn the collar on the tubing
clockwise until snug.
2. To remove the Motor or Prophy Handpiece, turn the collar on the tubing counter
clockwise until loose.
CYCLES:
The Motor and Prophy Handpiece can be subject to 250 cycles = (Performed
Sterilization Process) per EN ISO 5.10 before showing signs of deterioration.
SLOW SPEED MOTOR
& PROPHY HANDPIECE
1610110
1600102
1600005
SABLE 20K
SABLE 5K
SABLE
E-Type 4-Hole
E-Type 4-Hole
5K U-Type
Air Motor with
Air Motor
Hygienist
Internal Air/
Prophy Swivel
Water
Handpiece
Slow Speed Air
Slow Speed Air
Slow Speed Air
Driven Motor
Driven Motor
Driven Motor
ISO 7785-2
ISO 7785-2
U-Type Heads
Type and
Type and
Only
OEM type
OEM type
Up to 20,000
Up to 5000
Up to 5000
RPM
RPM
RPM
85 grams
110 grams
75 grams
Autoclave
Autoclave
Autoclave
134°C @8min,
132°C @10min,
132°C @10min,
dry time=20 min
dry time=40 min
dry time=40 min
250 + as per
250 + as per
250 + as per
ISO 7785-1
ISO 7785-1
ISO 7785-1
CLEANING: Preparation and Cleaning before Sterilization:
Clean and lubricate the handpiece as soon as possible after treatment. Failure to
properly maintain the handpiece may cause infection, overheating leading to burn
injuries.
DO NOT IMMERSE INSTRUMENTS IN DISINFECTANT SOLUTION
OR CLEAN IN ULTRASONIC DEVICES. MANUAL CLEANING ONLY.
1. PREPARATION:
a. Always wear protective gloves, a mask, and protective goggles for safety
purposes and to minimize the risk of infection.
b. Wipe the exterior of the handpiece/accessory using a cleaning wipe.
c. Operate the handpiece at the chair-side for at least 20 seconds to purge fluids
in the handpiece (not applicable for air motors).
d. Remove the handpiece, carry the handpiece to the decontamination area and
proceed to step 2.
2. CLEANING AND DRYING:
*Wash the handpiece and irrigation nozzle under running water following the procedures below.
(Water condition:<38°C (100°F), >3.5L/min; water should be same quality as drinking water)
a. Clean the external surfaces of the handpiece/accessory with a soft bristled
toothbrush for 15 seconds or more.
b. Under appropriate lighting (natural office lighting), inspect the handpiece/
accessory for blood or other debris. If any visible debris remain, repeat the
process until it is visually clean.
c. Wipe the exterior of the handpiece/accessory clean using a cloth moistened
with disinfectant or ethanol with a concentration of 60-90% (ie. Isopropyl).
LUBRICATION:
Recommended to use Sable Branded autoclavable handpiece lubricant.
1. Remove Motor or Prophy Handpiece from airline.
2. Using spray lubrication can, oil dropper or lubrication station, apply lubricant in
the drive air tube (smaller of the two tubes at the rear of the motor).
3. Reattach Motor or Prophy Handpiece to airline if not using a lubrication station.
4. Run entire tool at normal speed for 10 seconds.
5. Wipe off excess oil from entire tool surface.
AUTOCLAVING/STERILIZATION
IMPORTANT NOTE: This handpiece or motor do not come sterilized from
the manufacturer. It is also recommended that the handpiece or motor be
sterilized after each use. Sable Branded autoclaving or sterilization packaging
is recommended to preserve item in sterile condition. Please note that after
sterilization, handpiece should be used at your earliest convenience to make
certain that sterilization is maintained.
1. Clean before sterilization (see CLEANING above)
2. Lubricate before sterilization (see LUBRICATION above).
3. For part number 1600111 set parameters for Moist Heat Sterilizer. PreVac cycle
135 degrees Celsius for 3 minutes and 16 minute dry time.
4. For part numbers 1600110, 1610110 set parameters for Moist Heat Sterilizer.
PreVac cycle 134 degrees Celsius for 8 minutes and 20 minute dry time
5. For part numbers 1600005, 1600102 set parameters for Moist Heat Sterilizer.
PreVac cycle 132 degrees Celsius for 10 minutes and 40 minute dry time.
WARRANTY INFORMATION:
Sable Slow Speed Motor or Prophy Handpiece are warranted to be free of
manufacturing defects for a period of 12 months* from date of purchase for the
standard Motor or Prophy Handpiece.
*Warranty is conditional on proper use and maintenance of the Motor or Prophy
Handpiece. Warranty is limited to repair or replacement at manufacturers'
discretion. Under no circumstances will the manufacturer be held responsible
for incremental or consequential damages. Please contact point of purchase
authorized distributor for warranty repair.
DISCLAIMER:
Sable Industries Inc. recommends that the sterilization process outlined in the
Directions for use (DFU) described in this document be followed precisely. Sable
Industries Inc. is not responsible for any products supplied in which the end user
does not follow the DFU.
DEALER
Manufactured by
SABLE INDUSTRIES INC.
100 Campbell Ave. Suite 5
Kitchener, ON N2H 4X8, Canada
TEL: 1-800-368-8106
FAX: 1-519-579-9323
Email: info@sableindustriesinc.com
www.sableindustriesinc.com
PPIS1004 Rev-N

Publicité

loading

Sommaire des Matières pour Sable 1600111

  • Page 1 ISO 7785-1 ISO 7785-1 2. Lubricate before sterilization (see LUBRICATION above). 3. For part number 1600111 set parameters for Moist Heat Sterilizer. PreVac cycle SPECIAL PRECAUTIONS: 135 degrees Celsius for 3 minutes and 16 minute dry time. 1. Always exercise extreme caution when operating your Motor or Prophy 4.