Télécharger Imprimer la page

Operating, Shipping, And Storage Conditions; Accessories; Standard Icons; Calibration - iHealth PT3 Mode D'emploi

Publicité

Les langues disponibles

Les langues disponibles

Operating, shipping, and storage conditions

(1) Operating conditions
Range of operating environments
Temperature: 59℉-104℉(15℃- 40℃)
Humidity: 15-85%RH,non-condensing
Atmospheric Pressure: 70kPa~106kPa
(2) Transportation / storage conditions
Main Unit: -4℉~131℉(-20℃~ 55℃)
Humidity: 15-95%RH, non-condensing
Atmospheric Pressure: 70kPa~106kPa

Accessories

Use only accessories provided by the original manufacturer,
and check for any missing accessories.
1 IR Thermometer
1 Instruction manual

Standard icons

Attention: Follow instructions for use!
(The sign background color: blue. The sign
graphical symbol: white.)
The batteries and electronic instruments must
be disposed of in accordance
with the locally applicable regulation, not with
domestic waste.
Manufacturer information
Symbol for "WARNING"
Serial number
Authorized representative in the European
Community
IP code of the device: this device's grade of
against ingress of solid foreign objects -- ≥
12.5mm diameter (and the against access to
hazardous parts with nger); the grade of
waterproof is dripping (15° tilted).
0120
Complies with MDD 93/42/eec Requirements

Calibration

The thermometer is initially calibrated at the time of
manufacture. If this thermometer is used according to the
use instruction, periodic re-adjustment is not required. If any
time your question the accuracy of measurement, please
contact distributor or manufacturer, the contact information
is on the last page.
2 AAA batteries
1 Quick user guide
8

Storage

1) Do not put the thermometer under direct
sunlight, high temperature, or moist
environments. Do not allow it to come into
contact with re or harsh vibrations.
2) Take out the battery if the device is not used for
a long period of time.

Electromagnetic Compatibility (EMC) Tables

Guidance and manufacturer's declaration -
electromagnetic emissions
The PT3 device is intended for use in the electromagnetic
environment speci ed below.The customer or the user of the
PT3 should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 2
CISPR11
RF emissions
Class B
CISPR11
Harmonic
Not applicable
emissions
lEC61000-3-2
Voltage
Not applicable
uctuations/
ickeremissions
lEC61000-3-3
Guidance and manufacturer's declaration —
electromagnetic immunity
The PT3 device is intended for use in the electromagnetic
environment speci ed below.The customer or the user of the
PT3 should assure that it is used in such an environment.
Immunity test IEC 60601 Test
level
Electrostatic
±6kV contact
discharge(ESD)
±8kV air
lEC61000-4-2
Guidance and - manufacturer's declaration.
Electromagnetic immunity
The PT3 device is intended for use in the electromagnetic
environment speci ed below. The customer or the user of the
PT3 should assure that it is used in such an environment.
Electromagnetic environment
-guidance
The PT3 device must emit
electromagnetic energy in
order to perform its intended
function. Nearby electronic
equipment maybe a ected.
The PT3 device is suitable for
use in all establishments
other than domestic and
those directly connected to
the public low-voltage
power supply network that
supplies buildings used for
domestic
purposes.
Compliance
Electromagnetic
level
environment -
guidance
±6kV contact
Floors should be
±8kV air
wood , concrete or
ceramic tile.
If oors are
covered with
synthetic material,
the relative
humidity should
be at least 30%.
9

Publicité

loading

Produits Connexes pour iHealth PT3