Baxtran ARX Série Mode D'emploi page 11

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

PERCENTAGE DISPLAY F2.1 = 2
Display specification
Display [Pr 20.], representing 20.5 %.
Pr = current actual weight / range ×100%.
Press
to display switch in percentage and weight.
CHECK WEIGHT AND SELECTION SCALE FUNCTION F2.1 = 3
Function description
Set parameters such as F2.2 = A, F2.3 = B, F2.4 = C and F2.5 = D.
When display weight is X.
If X ≤ A, do not conduct check weight and selection.
If X < (B – D), it lacks of weight, and the display flickers.
If (B – D) ≤ X ≤ (B – C), it is qualified and the display has normal display
If X > (B – C), it is overweight and the display flickers.
Acquisition of target value
Press
long until the display shows [TAR ET], and then press
If F2.6 = 0, press
, the instrument will take the weight on current scale as the new target value and quit the setting interface.
If F2.6 = 1, the display shows [000000], to request manual change of target value. After change, press
setting interface.
FUNCTION OF ACCUMULATION SCALE F2.1 = 4
Operation method
In normal weighing state, when the scale is in , add weight to the scale and press
it indicates that the current display weight is included accumulated value, and then it returns to normal weighing state. If the display shows [--no--] for a
second, and returns to the normal weighing state, it indicates operation is invalid. Reason: 1. Between two accumulation operations, the scale
must have back-to- process, otherwise, accumulation is refused. 2. Accumulation operation is effective only when display weight≥F2.2 is set. 3. The
scale is in dynamic state.
Adjustment, clearing and printing of accumulated value
In normal weighing state, press
display shows current total accumulated value [A 9. 00] and flickers. To clear accumulated value, press [Pr 20.], to make flicker weight be 0.
Press
to print accumulated data. Press
Attention: set whether to be over detailed data or accumulated data in F4.6.
FUNCTION OF COUNTING SCALE F2.1 = 5
Instrument display
[c 128], showing current quantity.
Sampling method
1.
Check whether the scale is in , if not, press
2.
Place materials counted on the scale.
3. Press
long until the display shows [ AMPLE], and then press
counted just now, and press
on operation panel for more than 2 seconds, the display will show [Zero] for a second, and then the
to quit the interface.
for setting.
for confirmation. The instrument saves sampling data and quits the sampling interface. If F2.6 = 1, the
to show current target value and flicker.
on operation panel, if the display shows [Add--] progress bar,
. If F2.6 = 0, the display shows [PC 00]. Input the quantity
to save setting data and quit

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

ArxArx lcdArx lcd wide

Table des Matières