• The current will now slowly increase to its maximum of
2 mA, as shown in the app.
• If you at any point want to pause the stimulation,
press the pause button in the app or the button on the
headset.
• Follow the instructions in the app during the rest of the
session. The session will last for 30 minutes.
• When the time is up, the headset will automatically
start decreasing the current.
• Do not remove the headset before the app indicates
that the stimulation has stopped.
• The headset gives a "beep" signal when the
stimulation has finished.
• After the session, dispose of the Headset Pads and put
the Pad Holder Rings and headset back in the box.
• The headset will automatically turn itself off after a
few minutes of inactivity. Start it again by pressing the
button on the headset.
NOTE: Most users that experience reduced symptoms of
depression do so within 3-4 weeks. For a summary of safety &
clinical performance, see flowneuroscience.com
12. Charging the Headset
• Charge the headset if the LED indicator on the front of
the headset does not blink when pressing the button.
12 / ENGLISH
• Remove the charger port cover and charge using the
supplied Micro-USB charger and cable.
• The LED indicator blinks during charging and shows a
steady light when the headset is fully charged.
• Recharging a fully flat battery takes approximately an
hour.
• Be sure to close the charger port cover again after
charging.
13. Cleaning of the Headset
• It is recommended to clean the headset and especially
the electrodes after each use.
• To clean the device, ensure that the Micro-USB cable is
not connected to the headset and remove the pads.
• Use an IPA (isopropanol) wet wipe (70% is a good
choice).
• Clean the electrodes using a circular motion.
• Neither the device nor its accessories are supplied
sterilized nor require sterilization prior to use.
ENGLISH / 13