Télécharger Imprimer la page

Troubleshooting - VISIOMED TensioFlash KD-595 Manuel D'utilisation

Masquer les pouces Voir aussi pour TensioFlash KD-595:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
Weight:
Approx. 290 g (Apparatus)
Accessories:
Cuff (fit arm size 8.7~12.6in / 22~32cm), Instruction
Manual, 4 AA batteries
Operation Voltage
6V DC
*Specifications may be changed without notice in the event
of Improvement being made.

TROUBLESHOOTING

If you have trouble while using the unit, please check the
following points first.
ERROR DISPLAY
POSSIBLE CAUSE
No battery installation
Nothing is displayed when
Battery worn out
you push the POWER button
The polarities of batteries
placed wrongly
Check your cuff if any air
leakage
Upper arm cuff not properly
adjust or placement, hand or
Shows abnormal results -
body shaking.
LCD shows "EE"
Cuff too tight
Cuff too loose
Battery icon on
Battery low power
The hand with the cuff was
held lower than your heart
The systolic pressure
The cuff was not attached
value or diastolic
properly
pressure value too high
You moved your body or
spoke during measurement
The hand with the cuff was
The systolic pressure
held higher than your heart
value or diastolic
pressure value too low
You moved your body or
spoke during measurement.
Note: Relax for at least 10 minutes and then take another
measurement.
Visiomed - KD-595 - 032016
HOW TO CORRECT
Insert batteries
Replace new batteries
Insert battery in the correct
polarities
Replace upper arm cuff with
new one
Renew adjust upper arm
cuff position and tightness,
and keep hand still.
Re-measurement or send
back dealer for re-calibrate
pressure
Readjust cuff and measure
again
Replace battery, and
measure again
Keeping correct position and
gesture to measure again
GEBRUIKSAANWIJZING
0197
KD-595
Visiomed - KD-558 - Manuel d'utilisation - 032016
NL
31

Publicité

loading