Fitting The Electrode/Sensor - Mettler Toledo InTrac 787 Instructions D'utilisation

Masquer les pouces Voir aussi pour InTrac 787:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

All manuals and user guides at all-guides.com
InTrac
®
787
4.2.2

Fitting the electrode/sensor

1
3
5
1. Pull out the immersion tube.
2. Close the ball valve.
3. Disconnect the immersion tube right at the ball valve.
4. Disconnect the electrode holder by turning it counterclock-
wise.
5. Disconnect the electrode cable.
6. Remove the electrode.
7. Check washers and O-rings on electrodes for damage
and replace if necessary.
8. Remove watering cap from electrode tip and rinse electrode/
sensor with water.
9. Carefully insert the electrode into the electrode holder and
screw in by hand as far as it will go. Important: use no tools!
10. Re-Install the complete housing.
© 02/02 Mettler-Toledo GmbH, CH - 8606 Greifensee
Printed in Switzerland
2
4
6
InTrac 787-01-CH
11
7
52 402 481

Publicité

Table des Matières
loading

Table des Matières