Interface Sérielle - Ismatec MCP Process Mode D'emploi

Ismatec
Table des Matières

Publicité

Hinweis
Der Antrieb MCP Process hat beim
Einschalten eine Verzögerung von
3 Sekunden, bis die serielle
Schnittstelle zuverlässig anspricht.
Please note
When switching on the MCP
Process drive, the serial interface
will only respond after a delay of 3
seconds.
Remarque
Le moteur MCP Process a un retard
de 3 secondes lors de la mise en
route jusqu'à ce que l'interface
sérielle fonctionne de manière
fiable.
GND
TxD
Betriebserde
Sendedaten
Signal Ground
Transmitted Data
Prise de terre
Données
du système
envoyées
RxD
Empfangsdaten
Received Data
Données reçues
GND RxD TxD
5 4 3 2 1
Kommunikationsparameter
Communication parameters
Paramètres de communication
Standardeinstellung
Default setting
Réglage standard
9600 baud, 8 bit,
1 stopbit, no parity
(DIP Switch 1 off)
Alternative
1200 baud, 8 bit,
1 stopbit, no parity
(DIP Switch 1 on)
MCP Process/Ismatec SA/15.11.2000/CB/
GP
Serielle Schnittstelle
RS232 IN (Eingang, weiblich)
Der Anschluss erfolgt über eine
9-polige D-Buchse.
Adressierung
Jeder Befehl beginnt mit der
Pumpenadresse (1–8). Sie ist
werkseitig auf 1 eingestellt und
kann mit @ verändert werden.
Die Adresse wird permanent
gespeichert (d.h. auch nach
Ausschalten des Antriebs).
Die Adressierung ermöglicht die
Ansteuerung von bis zu 8
®
Ismatec
-Pumpen mit einer
RS232-Schnittstelle.
Aufbau der Befehle
Nach der Adresse folgt ein
Zeichen als Befehl. Manche
Befehle haben zusätzlich einen
Parameter, der aus 4 bzw. 5
Ziffern besteht.
Der Befehl wird mit einem
carriage return (ASCII 13) abge-
schlossen. Die meisten Befehle
quittiert die Pumpe mit einem
Stern *. Ja/Nein-Abfragen
beantwortet sie mit + (ja) oder –
(nein).
Mehrstellige Antworten
beschließt die Pumpe mit (cr)
ASCII 13 und (lf) ASCII 10.
Fehlerhafte Zeichensequenzen
quittiert sie mit einem #.
Zahlenwerte gibt die Pumpe als
drei- bis fünfstellige Zahl zurück.
Vier dieser fünf Stellen sind
Ziffern, eine Stelle ist entweder
ein Dezimalpunkt oder eine füh-
rende Leerstelle.
Serial interface
RS232 IN (female)
A 9-pin D-socket is available on
the rear panel of the pump.
Setting the pump address
Each command string begins
with the pump address (1–8).
It is factory-set at 1 and can be
changed with @.
The address will be stored
permanently (even after the
pump has been switched off).
The assignment of the address
enables the user to control up to
®
8 Ismatec
drives via one
interface.
Structure of the commands
The address is followed by a
character. Some commands have
an additional parameter which
always consists of 4 or 5 figures.
The command string is com-
pleted by a carriage return
(ASCII 13). The pump confirms
most of the commands with an
asterix *. Yes/No inquiries are
answered by + (yes) or – (no).
Multi-digit replies are concluded
by (cr) ASCII 13 and (lf) ASCII 10.
Incorrect command strings are
answered by #.
Numerical values are confirmed
as 3 to 5-digit figures. Four of the
five digits are numerals, one digit
is either a decimal point or a
preceding blank space.
Interface sérielle
RS232 IN (entrée femelle)
Le raccordement se fait par le
biais d'une douille D à 9 pôles
(femelle).
Adressage
Chaque commande commence
avec l'adresse de pompe 1–8. Elle
est réglée par défaut sur 1 et
peut être modifiée avec @.
L'adresse est enregistrée de
manière permanente (c'est-à-dire
également après la mise hors
service du moteur). L'adressage
permet la commande de plu-
®
sieurs moteurs Ismatec
par le
biais d'une seule et même
interface RS232.
Structure des commandes
Un signe suit l'adresse en tant
que commande. Certaines
commandes ont en outre un
paramètre comportant toujours
4, respectivement 5 chiffres.
La commande se termine par
un retour à la ligne (ASCII 13). La
pompe quittance la plupart des
commandes avec une étoile *.
Elle répond aux questions oui/
non avec + (oui) ou - (non).
La pompe termine les réponses
multipositions avec (cr) ASCII 13
et (If) ASCII 10.
Les séquences de signes
erronées sont quittancées avec
un #.
La pompe renvoie les valeurs
chiffrées sous forme de nombre
de trois à cinq positions. Quatre
de ces cinq positions sont des
chiffres, une position étant soit
un point décimal ou un blanc
(1
e
pos).
35

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ism 915

Table des Matières