ZVA-Z90E / -Z110E
Risk of instrument damage
The RF input power at the connectors RF IN and LO IN must not exceed the maximum
values quoted in the data sheet. The maximum values are below the maximum RF
source power of the network analyzer. The frequency converter mode ensures compat-
ible source powers.
Before you connect your converter to the network analyzer, always activate the fre-
quency converter mode using the "Frequency Converter" tab of the "System Configura-
tion" dialog (see
on page 20) and select the proper converter type and connecting diagram.
2.3.5 RF Connectors – Output
Two SMA connectors:
MEAS OUT (measurement signal output)
REF OUT (reference signal output)
For correct cabling, please refer to
on page 13.

2.3.6 Control Connector

The three-pin control connector receives the control signal for the source power of the
frequency converter.
For correct connection read
on page 13.

2.4 Putting the Converter into Operation

This section describes the basic steps to be taken when setting up the frequency con-
verter for the first time.
Getting Started 1307.7197.62 ─ 08
Chapter 3.3, "Activating the Frequency Converter
Chapter 2.4.6, "Connecting RF
Chapter 2.4.5, "Connecting the Control
Preparing for Use
Putting the Converter into Operation
Mode",
Cables",
Cable",