5. Symbols
YYMMxxxx
6. Safety instructions
-
To avoid any risk of fire or explosion:
Do not smoke in the vicinity of the device.
Do not use near an ignition point.
Do not place the device in contact with greasy products.
Do not handle the device with greasy hands.
Do not dismantle a pressurised appliance.
-
Do not restrict the output of the medical device.
-
Influence of pressure: Caution, check the supply pressure to the device. If the inlet pressure is not within the range
indicated on the unit, the outflow could waiver from the standard authorised limits.
-
Influence of temperature: Please note that temperature variations (between 0°C and 40°C) influence the accuracy of
the outflow delivered.
-
Do not disassemble the device's inlet or outlet connectors.
-
Ensure that the device is always in a vertical position.
-
Warning! Do not allow water to enter the flow meter during transport with a humidifier; always unscrew the humidifier
before handling.
7. Cleaning and disinfection
-
Disconnect the appliance from the gas supply. Clean the outside of the unit with a clean, damp cloth. Allow the device
to dry. If using decontaminants, check their compatibility with the plastic (see technical sheet).
-
Do not immerse.
-
Do not use surface decontaminants.
8. Storage
Store at between -20 and 60°C in a clean, dry place.
If possible, preserve the packaging.
9. Maintenance
RTM3 requires no preventive maintenance during its lifetime.
The device can only be disassembled by a person authorised to do so by Technologie Médicale.
Technologie Médicale recommends an annual inspection of the device:
Check the connection to the wall outlet or the pressure regulator outlet.
Check that the gas is released correctly at different rates.
Check that there is no leakage when the valve is closed.
10. Guarantee
Device guaranteed for 1 year including parts and labour, except for damage or accidents resulting from negligence,
inappropriate use, and lack of supervision or maintenance.
Contractual life: 10 years
In accordance with European Directive 93/42/EEC of 14 June 1993, any incident
or risk of incident must be reported to TECHNOLOGIE MEDICALE immediately.
Do not use grease, lubricant or oil
See instructions
Serial number engraved on the device:
YY: Year of manufacture, MM: Month of manufacture, xxxx: Unit number