solution for a more efficient distribution of the medication on the nasal mucous
membranes.
12. TOnce the application is complete, switch off the device and clean it, following the
instructions provided in the "CLEANING, SANITISATION, DISINFECTION" paragraph.
CLEANING SANITISATION DISINFECTION
Switch off the device before any cleaning procedure and unplug the power cable from
the socket.
DEVICE AND TUBING EXTERIOR
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of
any sort).
ACCESSORI
To disassemble the bowl, firmly grasp the bowl body
and then use your other hand to rotate the particle size
regulator (C1.1) clockwise until it is ejected from the up-
per part (C1.2) as a result of unscrewing. The selector
(C1.1c), valve (C1.1b), regulator (C1.1a) and upper
part (C1.2) can then be disassembled to be cleaned.
The nozzle (C1.3) is interference fitted in the upper part
(C1.2) and can be extracted by simply pulling it out-
wards.
Disassemble the NASAL WASH as illustrated in the connection
diagram. As the separator (C6.2) is fitted in the body (C6.4),
it is necessary to exert some force to separate them.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebulizer cup and the
accessories, choosing one of the methods described below.
method A: sanitise accessories C1.1- C1.1a-C1.1b-C1.1c-C1.2-
C1.3-C1.4-C2-C3-C4-C5-C6.1-C6.2-C6.3-C6.4 under potable
hot water (approximately 40 °C) with a gentle, non abrasive
dish detergent.
method B: sanitise accessories C1.1- C1.1a-C1.1b-C1.1c-C1.2-C1.3-C1.4-C2-C3-C4-C5-
C6.1-C6.2-C6.3-C6.4 in the dishwasher with a hot cycle.
method C: sanitise accessories C1.1- C1.1a-C1.1b-C1.1c-C1.2-C1.3-C1.4-C2-C3-C4-C5-
C6.1-C6.2-C6.3-C6.4 by immersing them in a solution of 50% water and 50% white
vinegar, then rinse thoroughly under potable hot water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After sanitising the accessories, shake them well without scrubbing them with clean
cloths and place them on a paper towel. Alternatively, dry them with a jet of hot air (for
example a hair dryer).
DISINFECTION
After sanitising the nebulizer cup and the accessories, disinfect them choosing one of
the methods described below.
method A: Accessories can be disinfected are C1.1- C1.1a-C1.1b-C1.1c-C1.2-C1.3-C1.4-
C2-C3-C4-C5-C6.1-C6.2-C6.3-C6.4 .
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlo-
rite) specific for disinfecting, which is available in any pharmacy.