Servicing And Maintenance - Chopin EM10 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN
4.
Servicing and
maintenance
30
EM10 Multi compartment oven
This apparatus does not require any particular maintenance.
Cleaning
Once a year, clean the apparatus using a slightly damp cloth.
Clean the inside of the cells with compressed air.
Temperature check
Each time the temperature setpoint is changed, and at least
every six months, check the measured temperature at the level of
the superior central screw with a precision thermometer ± 1 °C.
If there is a temperature difference during this operation, contact
your CHOPIN Technologies representative.
Fuse replacement
If the apparatus does not switch on and if the power supply has
been checked, replace the EM10 fuse.
Disconnect the apparatus and replace its fuse by using a small
screwdriver.
The fuse is located at the back of the apparatus behind a cover.
Fuse characteristics: 5x20 T 2A 250V.
User's manual
12/2016

Publicité

Table des Matières
loading

Table des Matières