Unit Switch between Fahrenheit and Celsius
To change LCD display from °C to °F: When the unit in Power off condition, press and hold Scan
button, do not release, and press ON/OFF button and hold, until the LCD Display show '°F'.
The same process can change LCD display from °F to °C.

Memory Function

If the reading of the thermometer is within the normal temperature range of 34°C to 42.2°C
(93.2°F to 108.0°F), the data is saved into memory each time when power is turned off.
** Press the "ON/OFF Memory" button again to see the temperature stored.

Changing the Battery

This device is supplied with one lithium cell (CR2032 x 1). Remember that Power off the unit before
replacing the battery. If power is on when the battery is replaced, a malfunction may occur. If a
malfunction occurs, restart the device. Keep the battery away from children.
Specifications:
**Storage Range: -20°C~50°C (-4°F~122°F) 15% ~ 85% RH
Compliance with ASTM E1965-98, Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/
EN60601-1-2(EMC), IEC/EN60601-1 (Safety) standards, ISO10993, RoHS.
Operating temperature range: 10~40°C (50~104°F) RH
85%
Transport conditions < 70°C, RH
95%
a. The "Clinical Bias" is -0.2 ~ -0.4°C.
b. The "Limits of Agreement" is 0.58.
c. The 'Repeatability" is 0.17°C.
Accuracy:+/-0.2 °C (0.4 °F) during 35.5~42°C (95.9~107.6°F), +/-0.3 °C (0.5 °F) for other range.
Atmospheric pressure: 800-1013hPa
Dimensions: 141.4 x 40.4 x 43.5mm
Weight: 67.3g.
Battery Life: Approx. 3000 times measurement