HOLD FUNCTION: In case the indication isn't stable or the reading is laboured, HOLD function will let you
memorize the weight displayed. Pressing HOLD/UNIT key, the indication will be locked and the HOLD indicator
will be ON. By pressing HOLD/UNIT key again, HOLD indicator will turn OFF and the device wil, return to
standard weighing mode.
ADVANCED OPERATION: Turn-OFF the device. Turn it ON while pressing ZERO/TARE key  display shows "
CAL
CALIBRATION: Once "
Remove any loads from the hook and press ZERO/TARE key again. The display shows the calibration weight and
by pressing HOLD/UNIT key it's possibile to change the value.
Once selected the desired calibration weight, press ZERO/TARE key and the display shows "
Place the selected weight shown on the display and press ZERO/TARE key again;
When the display shows "
SETTING CAPACITY: Once display shows "
Press ZERO/TARE key to enter capacity mode and the display shows the current capacity. Press HOLD/UNIT key
to select the desired capacity. Press ZERO/TARE key again to confirm.
INTERNAL COUNTS: Once display shows "
Press ZERO/TARE key to show internal counts.
SET AUTO POWER OFF: Once display shows "
Press ZERO/TARE key to enter settings, than press HOLD/UNIT key to select auto power-off time:
oF 0
: disabled
oF 3
: auto-OFF after 3 minutes of stand-by
oF 10
: auto-OFF after 10 minutes stand-by
Press ZERO/TARE key to confirm.
The instrument must be cleaned with a soft cloth, dampened with water or neutral detergent, avoiding the use of
solvents or abrasive substances. Do not use large amounts of water while cleaning it, as it may cause damage to the
electrical components. Always remove batteries before cleaning. In case of prolonged non-use of the instrument,
remove the batteries from the terminal and cover the instrumentation to keep it intact. During transport, be careful
not to subject the instrument to shocks or excessive mechanical stress. In case of repair or assistance, contact your
dealer or an authorized service center by contacting
DYNAMOMETER CR REV. 01 del 13/12/2018
" is shown on the display, press ZERO/TARE key and the display shows "
PASS
" it means that calibration is completed.
CAL
", press HOLD/UNIT key and the display shows "
CAL
", press HOLD/UNIT key 2 times and the display shows "
CAL

5. CLEANING AND MAINTENANCE

" press HOLD/UNIT key3 times and the display shows "
service2@wunder.it
or
sales@wunder.it
CAL
"
UnLoAd
".
LoAd
".
CAP
".
inP
".
oFF
".