Step 4a: Pull-Off Test
Step 4a: Pull-Off Test
Actuator Handle
Hose
Actuator
Assembly
Quick Coupling
The PosiTest AT-M powers-up and displays dashes when the
pressed. To preserve battery life, the instrument powers down after 5 minutes of
no activity.
Quick Guide
(1) Open the pressure relief valve completely (turn counter clockwise)
(2) Connect the actuator to the dolly
(3) Close the pressure relief valve completely (turn clockwise)
(4) Zero - Press the
(5) Pump pressure into the system until the dolly pulls the coating away
1.
Ensure the pressure relief valve on the pump is completely open. (turn
counter clockwise)
2.
Push the actuator handle completely down into the actuator assembly. Place
the actuator assembly over the dolly head and attach the quick coupling to the
dolly by reaching through the holes in the actuator assembly and lifting the quick
coupling. Release the quick coupling when the dolly head is completely engaged.
(PosiTest AT-M Manual)
LCD
USB Data Transfer /
Power Port
Zero button.
OPEN
Page 6
Pump Handle
Pump
CLOSE
button is