EIJKELKAMP 18.52.01 Manuel D'instructions page 23

Masquer les pouces Voir aussi pour 18.52.01:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Conductivity
Cell Constant
Normalization
Conductivity
Temperature
Linear
Conductivity
Temperature
Coeffcient
Pure Water
Conductivity
Coefficient
DO/ Salinity
Set salinity
Auto Salinity
DO/ Salinity
compensation
The cell constant can be adjusted from 0.010 to 10.000
To select a normalized temperature that will be used to
normalize conductivity measurements to a standard temperature.
Factory default is 25 °C
This value reflects the degree to which the conductivity is
affected by temperature changes. Factory deafault is 2.1% per
°C
Pure water coefficient will be calculated and applied
automatically for ultra pure water measurement if 'ENABLE' at
appropriate set up pages
Sets the salinity (in ppt) for your solution. By setting the salinity,
the meter will compensate for the impact of salinity on the partial
pressure/oxygen concentration relationship of your sample.
Range: 0 to 50 ppt.
If enabled, the meter automatically compensates for salinity.
Auto salinity compensation is applicable only in multi
measurement mode and only if one of the measurements
selected is conductivity related. This option is activated only in
CD650 & PCD650. Otherwise, manually set salinity value will be
applied.
23
2.7.1
2.7.1
2.7.1
2.7.1
2.8.2
2.8.2

Publicité

Table des Matières
loading

Produits Connexes pour EIJKELKAMP 18.52.01

Ce manuel est également adapté pour:

18.50.01

Table des Matières