Troubleshooting Guide

Most problems can be avoided by reading and following the instructions in this operating manual. Below we list some of those
most commonly experienced along with suggestions for solving them. If, however, these should not resolve the issue, or if you have
questions not covered below, please contact Fisher Scientific.
Trouble
Air bubbles appear in discharge nozzle
Barrel does not fill with liquid
Dispensing not possible
Wrong dispense volume
Possible cause
Liquid reservoir is empty
Too fast filling action
Leaking piston
Leaking discharge valve
Inlet tube not fitted correctly
Blocked dispense nozzle
Discharge valve stuck
Instrument not calibrated
www.eu.fishersci.com/fisherbrand
Solution
Refill reservoir and prime unit
Fill and dispense more slowly
Clean piston. If problem persists
replace piston
Clean unit by flushing throughly.
If problem persists, replace the
discharge valve
Connect inlet tube correctly
Disassemble the dispense nozzle
and flush with cleaning fluid
Clean unit by immersing valve
assembly in cleaning fluid. If
problem persists, replace valve
assembly
Follow steps of User Calibration
14