Maintenance

Element of Oil Mist Trap are subject to replacement every 1,500 to
2,000 hour of operation due to filling up of the element by dust.
Whole gasket also should be replaced at the same time.
For part replacement is available;
§
This part is available at our local AGILENT service.

Cartridge changing

The cartridge has a limited life time that depends on the pump
application. The cartridge clogging activates the by-pass device so
that the gases from the pump are directly sent to the external
ambient to avoid dangerous pump overpressures.
In this case the oil mist are diffused into the atmosphere indicating
that the cartridge is obstructed and, to avoid the ambient pollution, it
is advisable to change the cartridge.

Retention

The Oil Mist Trap or any other component in a bag (such as a plastic
bag). Keep it away from the dust and the direct sunshine.