n C160 : 1000 A calibre (2000 A max. peak)
Intensity in A~
Intrinsic error
(1)
Dephasing
C160 : 100 A calibre (300 A max. peak)
Intensity in A~
Intrinsic error
(1)
Dephasing
C160 : 10 A calibre (30 A max. peak)
Intensity in A~
Intrinsic error
(1)
Dephasing
(1) As % of output signal
n C173 : 1000 A calibre
Intensity in A~
Intrinsic error
(1)
Dephasing
C173 : 100 A calibre
Intensity in A~
Intrinsic error
(1)
Dephasing
C173 : 10 A calibre
Intensity in A~
Intrinsic error
(1)
Dephasing
C173 : 1 A calibre
Intensity in A~
Intrinsic error
(1)
Dephasing
(1) As % of output signal
4.4
Conditions of use
The "C100" clamps must be used in the following conditions, in order to comply with user safety and
metrological performance requirements.
ICS Schneider Messtechnik GmbH
Briesestraße 59
D-16562 Hohen Neuendorf / O T Bergfelde
1...50 A
50...200 A
≤ 1% + 1 mV
≤ 3°
(2)
0.1...5 A
5...20 A
≤ 2% + 5 mV
≤ 15°
(2)
0.1...0.5 A
0.5...2 A
≤ 3% + 10 mV
(2)
(2)
(2) Unspecified
1...10 A
10...100 A
≤ 1% + 0.2 mV
≤ 0,5% + 0.2 mV
≤ 2°
(2)
0.1...1 A
1...10 A
≤ 1% + 0.2 mV
≤ 0.5% + 0.2 mV
≤ 2°
(2)
0.01...0.1 A
0.1...1 A
≤ 1% + 0.2 mV
≤ 0.5% + 0.2 mV
≤ 5°
(2)
0.001...0.01 A
0.01...0.1 A
≤ 3% + 1 mV
≤ 3% + 1 mV
(2)
(2)
(2) Unspecified
13
Tel.: 03303 / 504066
Fax: 03303 / 504068
200...1000 A
1000...1200 A
≤ 2°
≤ 1°
20...100 A
100...120 A
≤ 10°
≤ 5°
2...10 A
10...12 A
≤ 15°
≤ 15°
100...1000 A
1000...1200 A
≤ 0.2%
≤ 0.2%
≤ 1°
≤ 1°
10...100 A
100...120 A
≤ 0.3%
≤ 0.2%
≤ 1°
≤ 1°
1...10 A
10...12 A
≤ 0.5%
≤ 0.5%
≤ 2°
≤ 2°
0.1...1 A
1...1.2 A
≤ 0.7% + 1 mV
≤ 0.7% + 1 mV
≤ 10°
≤ 10°
info@ics-schneide r.de
www.ics-schneide r.de