■ It is recommended to fill the reservoirs only with the amount of liquid needed for the working day.
®
X-Cidol
2 should be refilled every day.
®
1 litre of X-Cidol
2 and 1 litre of water are sufficient for 13 cleaning cycles with 3 selected instrument stations.
Water should be refilled on a weekly basis.

3) Lubricant reservoir:

The oil reservoir is positioned on the side of the device (photo 1). When the oil amount is insufficient, the device
displays an error indication (see paragraph malfunction displays).
In order to refill the reservoir, unscrew the cover and pour the TSU oil bottle's content into the reservoir, ref
50700024 (photo 2).
Before moving the device into a horizontal position, close the valve in the middle of the cover by pushing it from the
screwing side (photo 3). The valve exceeds the cover when it is completely closed (photo 4).