A calibration has to be repeated
• after changing/cleaning of the strip adapter
and/or the internal standard
• in the case of suspect measurement results during the
boot up (self-check test failed)
• in the case of error messages E07 or E10
(after cleaning the optics)
• in the case of RQcheck failure (error messages E09
and E11)
• in the case of suspect measurement results
• after severe mechanical distress (e. g. after dropping the
instrument).
3

Measurement

There are two scenarios on how to start a measurement:
1. The desired method is not yet saved in the instrument.
Continue with chapter 3.1.
2. The method is already saved in the instrument.
Continue with chapter 3.2.

3.1 Add a new method

There are two scenarios for adding a method:
1. Guided way
2. Shortcut