7
Annexe C : « Données de correction personnalisées » types
Personal wavelength correction
Spectral Absorption Plot measured for: THZ12D-3S-VP-D0 Power Detector
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
0
100
Guide d'utilisation des appareils de série THZ-D
Révision 1.4
200
300
400
Wavelength (µm)
TM
Certificate
Personal Wavelength Correction
(µm)
* 10.6
43
52
70
96
119
122
134
158
184
191
214
237
268
288
334
349
394
440
600
* Calibration w avelength
For Gentec-EO monitors, select the proper
500
600
w avelength in menu
For other monitors, multiply by the correction multiplier
Pow er corrected = Pow er read x correction multiplier
Example: Pow er (158µm) = 10mW x 1.897 = 18.97 mW
Gentec Electro-Optics Inc. Tous droits réservés
Serial #Typical
Wavelength
Correction
(THz)
multiplier
* 28.3
6.97
5.77
4.28
3.12
2.52
2.46
2.24
1.90
1.63
1.57
1.40
1.26
1.12
1.04
0.898
0.859
0.761
0.681
0.500
20
TM
1.000
0.945
0.938
0.942
0.963
0.978
0.980
0.982
0.996
1.005
1.007
1.013
1.016
1.019
1.021
1.028
1.032
1.026
1.026
1.025