Warnings and Cautions

CAUTION: Repeated dry firing will damage
Injector. Dry firing is activating the Injector
with either an empty Syringe or without a
Syringe.
CAUTION: The Injector may be damaged
if not fully nested in the Reset Station when
closing the lid.
WARNING: The tip of the Filling/Vial
Adapter is sharp. Use caution when handling.
WARNING: Syringes auto-disable during
use. Attempting to use a Syringe more than
one time may result in contamination or
device malfunction.
WARNING: Vial integrity or sterility is
not ensured when used with the Filling/
Vial Adapter. DO NOT reuse the Filling/
Vial Adapter in another vial. Doing so may
contaminate contents of new vial.
CAUTION: The protective Syringe
Cap is designed to protect against touch
contamination only, and does not provide a
barrier against other forms of contamination.
CAUTION: Visually inspect sterile package
for damage. DO NOT use if torn or damaged.
DO NOT re-sterilize.
CAUTION: DO NOT attempt to use a
partially filled Syringe; damage may occur.
WARNING: DO NOT direct the Syringe
toward the face or nearby skin when loading
the Syringe into an Injector, preparing to give
an injection, or ejecting a Syringe after use.
Injury may occur.
CAUTION: Partial insertion of a Syringe
into an Injector may damage the Syringe
and/or Injector.
CAUTION: DO NOT depress the Multi-
function Button if the Polished Core Ring is
not visible as shown in figure 10. This may
cause inadvertent activation.
1) Prepare Injector
2) Fill Syringe
3) Load Injector