Care and Sterilization Instructions for Model 1140M Flex Handle, Model 1173B/R
Sizers, and Model SET1173 Tray
Instructions d'entretien et de stérilisation de la poignée flexible, modèle 1140M,
des calibreurs, modèles 1173B/R et du plateau, modèle SET1173
English (EN)...............................................................................................................................................................................................................................1
Français (FR)..............................................................................................................................................................................................................................4
Symbol Legend ■ Légende des symboles.....................................................................................................................................................................7
English
1.0 Accessories
These instructions apply to the following accessories: model 1140M handle, model 1173B/R sizers, and model SET1173 tray
and lid, for use with MITRIS RESILIA mitral valve, model 11400M.
The Instructions For Use supplied with MITRIS RESILIA mitral valve, model 11400M provide information on the use of these
accessories.
2.0 How Supplied
The accessories are packaged separately, provided non-sterile and must be cleaned, disinfected and sterilized before
each use. They cannot be sterilized in their original packaging.
3.0 Cleaning Instructions
The accessories must be cleaned, disinfected, and sterilized prior to each use in accordance with appropriate hospital
procedures and the following recommended instructions and parameters.
The accessories should be retired from use when signs of cracking, crazing, or other deterioration is evident. Contact your
Edwards Lifesciences sales representative to obtain appropriate replacements.
CAUTION: Examine the accessories for signs of wear, such as dullness, cracking or crazing. Replace the handled if any
deterioration is observed.
WARNING: Fragments of handles and sizers are not radiopaque and cannot be located by means of an external
imaging device.
3.1 Instructions for Manual Cleaning:
Step
Procedure
1
Rinse the handle under cold, running tap water to remove gross soil.
2
Prepare a neutral pH cleaning solution (such as Enzol or equivalent) per the manufacturer's
recommendation (1 oz per 1 gallon (30 ml per 3.8 liters) of warm tap water).
3
Fully immerse the handle in the cleaning solution and soak for a minimum of 5 minutes. Avoid contact
between the accessories.
Edwards, Edwards Lifesciences, the stylized E logo, MITRIS, MITRIS RESILIA, and RESILIA are trademarks of
Edwards Lifesciences Corporation. All other trademarks are the property of their respective owners.
Directory ■ Annuaire
1

Publicité

loading

Sommaire des Matières pour Edwards 1140M

  • Page 1 English 1.0 Accessories These instructions apply to the following accessories: model 1140M handle, model 1173B/R sizers, and model SET1173 tray and lid, for use with MITRIS RESILIA mitral valve, model 11400M. The Instructions For Use supplied with MITRIS RESILIA mitral valve, model 11400M provide information on the use of these accessories.
  • Page 2 Step Procedure After the 5-minute soak time, use a soft-bristled brush and scrub the accessories for a minimum of 3 minutes to remove any visible soil. Remove the accessories from the cleaning solution and rinse with cold, running tap water. Then rinse in reverse osmosis/deionized water for a minimum of 2 minutes.
  • Page 3 4.2 Instructions for High Level Disinfection: Step Procedure Equilibrate a bath of high level disinfectant (Cidex OPA or equivalent) to a minimum of 20 °C (68 °F), or follow manufacturer's recommendations. Fully immerse accessories in the disinfectant solution and ensure all air bubbles are removed from the surface by wiping with a sterile lint-free cloth.
  • Page 4 Français 1.0 Accessoires Ces instructions s’appliquent aux accessoires suivants : poignée, modèle 1140M, calibreurs, modèles 1173B/R et plateau et couvercle, modèle SET1173, pour l’utilisation de la valve mitrale MITRIS RESILIA, modèle 11400M. Le mode d’emploi livré avec la valve mitrale MITRIS RESILIA, modèle 11400M, fournit des informations sur l’utilisation de ces accessoires.
  • Page 5 Étape Procédure Insérer le plateau chargé dans le laveur automatique avec le couvercle détaché. Si possible, régler le moteur du laveur sur une vitesse ÉLEVÉE et sélectionner le cycle et les paramètres conformément aux recommandations suivantes : PHASE TEMPS DE RECIRCULATION TEMPÉRATURE TYPE DE DÉTERGENT (MINUTES)
  • Page 6 AVERTISSEMENT : ne pas empiler les plateaux pendant la stérilisation. L’utilisateur est responsable de la qualification des écarts par rapport à la méthode de nettoyage ou de stérilisation recommandée. Type de Configuration Précondi‐ Température Température Temps Temps de Temps de stérilisateur tionnement minimale...
  • Page 7 Symbol Legend ■ Légende des symboles English Français English Français Reg. Reg. No.¹ No.¹ Consult Catalogue Référence Consulter le mode 2493 instructions for Number catalogue 1641 d’emploi sur le use on the site Web website Quantity Quantité 0434A Caution Avertissement 2492 Lot Number N°...
  • Page 8 06/21 10048310001 A © Copyright 2021, Edwards Lifesciences LLC All rights reserved. Telephone 949.250.2500 Web IFU 800.424.3278 Edwards Lifesciences LLC 949.250.2525 One Edwards Way Irvine, CA 92614 USA...

Ce manuel est également adapté pour:

1173b/rSet1173