Télécharger Imprimer la page

Publicité

Liens rapides

Rated Max: 125VAC, 60Hz, 12 Amp Regulatory Derated
Non-Surge
5302M-1N
Surge
5302M-1S
Surge with Alarm
53A2M-1S
WARNINGS AND CAUTIONS:
• TO AVOID FIRE, SHOCK, OR DEATH; TURN OFF POWER at circuit breaker or fuse
and test that power is off before wiring!
• FOR INDOOR USE ONLY.
• To be installed and/or used in accordance with appropriate electrical codes and
regulations.
• If you are not sure about any part of these instructions, consult an electrician.
• Electromagnetic field may interfere with other equipment. If interference occurs, move
medical strip away from equipment being affected.
• Only connect to a properly grounded outlet.
• Do not plug into another re-locatable power tap.
DESCRIPTION
Leviton's Medical Grade Power Strips are designed for use in rugged medical environments
such as patient care areas. They comply with UL 60601-1 (Medical Electrical Equipment)
and UL 60950-1 (Information Technology Equipment) standards and are certified to UL
1363A (Special Purpose Re-locatable Power Tap) to supply power to plug-connected
medical equipment in patient care vicinities as defined in NFPA 99. These power strips
have ETL determined compliance to all UL and CSA standards mentioned in this instruction
sheet.
These Medical Grade Power Strips are available in non-surge and surge models. Surge
models meet the UL 1449 3rd Edition standard and are classified as Type 3 surge protective
devices. The strips are available in 2, 4, or 6 outlet versions with 7 or 15 foot power cords.
Each outlet is protected by a unique cover designed specifically for patient care areas.
For a list of all standards and compliance see specifications section below.
SPECIFICATIONS
All Models
• Limited Lifetime Warranty
• NFPA 99 - Complies with requirements
as it pertains to patient care vicinities
• UL 60601-1, UL 60950-1, UL 1363A
• CSA – C22.2 (No. 601.1-M90,
No. 60950-1-07, & No. 21)
• NOM 057
• Hospital Grade receptacles
• Hospital Grade right angle molded plug
• Lockable individual outlet covers
• Resettable, 15 or 20 Amp circuit breaker
• Rugged metal housing
• Easily accommodates surface mounting
• Green power LED indicates power is
present
• Class 1 equipment
• Operating temperature: 0
C to 40
o
• Relative humidity: 5% to 95% non-
condensing
• Transport and storage temperature: -10
OPERATION
1. Plug power cord into 125VAC outlet.
2. All Models: The green "POWER" LED will illuminate when power is applied.
Surge Models only: The green "PROTECTION" LED will illuminate when power is
applied.
3. To open outlet cover, use a plug blade of the component to be plugged in as shown in
FIGURE 1.
4. To close cover, press down until it snaps closed.
5. Operating symbols:
- Power LED: Indicates power is present when lit.
- Protection LED (surge models only): Indicates surge protection is active when lit.
-
: Green dot indicates hospital grade outlet.
-
: AC Symbol indicates alternating current.
-
: Earth ground symbol indicates grounding electrical point.
-
: Meets direct support requirements for current carrying parts at 120V rms or less
and 15A or less.
INSTALLATION - FLYING LEADS MODELS ONLY
Models containing flying leads (no power plug) must be hardwired correctly to maintain
power, protection and polarity status.
1. WARNING: TO AVOID FIRE, SHOCK, OR DEATH; TURN OFF POWER at circuit
breaker or fuse and test that power is off before wiring!
2. Connect leads to 125VAC single phase power source as shown in FIGURE 2.
3. Restore power at circuit breaker or fuse.
MOUNTING
Medical strips can be surface mounted using mounting holes on either end of strip.
1. Mark mounting surface using medical strip as template.
2. Drill holes and attach strip using screws (not supplied).
3. Refer to FIGURE 3 for mounting dimensions.
CARE AND MAINTENANCE
WARNING: Do not spray with liquid cleaning products.
To clean, unplug unit and wipe with a clean cloth.
Check status LED's monthly. Normal operation requires LED's to be lit.
Check circuit breaker (CB) status monthly. Normal operation requires CB to be depressed.
Leviton warrants to the original consumer purchaser and not for the benefit of anyone else that this product at the time of its sale by Leviton is free of defects in materials and workmanship under normal
and proper use during the lifetime of the product. Leviton's only obligation is to correct such defects by repair or replacement, at its option, if the product is returned prepaid, with proof of purchase
date, and a description of the problem to Leviton Manufacturing Co., Inc., Att: Quality Assurance Department, 201 North Service Road, Melville, N.Y. 11747. This warranty excludes and there
is disclaimed liability for labor for removal of this product or reinstallation. This warranty is void if this product is installed improperly or in an improper environment, overloaded, misused, opened,
abused, or altered in any manner, or is not used under normal operating conditions or not in accordance with any labels or instructions. There are no other or implied warranties of any kind, including
merchantability and fitness for a particular purpose. Leviton is not liable for incidental, indirect, special, or consequential damages, including without limitation, damage to, or loss of use of, any equipment,
lost sales or profits or delay or failure to perform this warranty obligation. The remedies provided herein are the exclusive remedies under this warranty, whether based on contract, tort or otherwise.
FOR CANADA ONLY:
For warranty information and/or product returns, residents of Canada should contact Leviton in writing at Leviton Manufacturing of Canada Ltd to the attention of the Quality Assurance Department, 165
Hymus Blvd, Pointe-Claire (Quebec), Canada H9R 1E9 or by telephone at 1 800 405-5320.
Medical Grade Power Strips
5304M-1N
5304M-1S
53A4M-1S
Surge Models Only
• UL 1449 3rd Ed. Type 3 device
• CSA – C22.2 No. 8
• 18mm MOV fused SPD circuitry
• Green protection LED indicates surge
protection is active
• Multi-line protection (L-N, N-G, L-G)
• Voltage Protection Ratings (VPR):
L-N:600Vpk
L-G: 600Vpk
N-G: 500Vpk
• Maximum surge current rating per mode
(L-N): 27KA
• Joule rating: 644j
• Maximum Continous Operating Voltage
(MCOV): 150V
• EMI/RFI noise reduction:
C
-6db to -29db (100KHz to 100MHz)
o
C to +50
C
o
o
LIMITED LIFETIME WARRANTY AND EXCLUSIONS
For Technical Assistance Call: 1-800-824-3005 (U.S.A. Only) www.leviton.com
Rated Max: 125VAC, 60Hz, 16 Amp Regulatory Derated
5306M-1N
Non-Surge
5306M-1S
Surge
53A6M-1S
Surge with Alarm
INSTALLATION
WARNINGS AND CAUTIONS:
• No user serviceable parts inside.
• Only 2 and 4 outlet Medical Grade power strips can be used with emergency trolley carts.
SURGE MODELS ONLY:
• Do not install this device if there is not at least 30 feet (10 meters) or more of wire
between the electrical outlet and the electrical service panel.
• No user serviceable parts inside. This device features internal protection that will
disconnect the surge protective component at the end of its useful life but will maintain
power to the load now unprotected. If this situation is undesirable for the application,
follow the manufacturer's instructions for replacing the device.
IMPORTANT SAFETY INSTRUCTIONS
SAVE THESE INSTRUCTIONS
1
Insert plug blade to release cover.
Insérer la lame d'une fiche pour libérer le
capuchon.
Inserte la cuchilla de la clavija para sacar
la cubierta.
16.87 in
(42.84 cm)
15.00 in
(38.10 cm)
(Top View | Vue du dessus
Vista superior)
5302M-2N
5302M-2S
53A2M-2S
FIGURE 1 | FIGURE 1 | FIGURA 1
2
Press on edge of cover to snap closed.
Appuyer sur le rebord du capuchon pour
l'enclencher.
Presione en el borde la cubierta para
cerrar.
FIGURE 2 | FIGURE 2 | FIGURA 2
Brown |Brun|Marrón
Blue |Bleu |Azul
Green/Yellow |Vert / Jaune
Verde / Amarillo
FIGURE 3 | FIGURE 3 | FIGURA 3
DIMENSIONAL SPECIFICATIONS | SCHÉMAS DIMENSIONNELS
ESPECIFICACIONES DIMENSIONALES
MEDICAL GRADE
COMPLIANT WITH
NFPA 99 REQUIREMENTS
FOR PATIENT CARE AREAS
Power
Protection
11.93 in
(30.30 cm)
MEDICAL GRADE
COMPLIANT WITH
NFPA 99 REQUIREMENTS
1
2
Insert plug blade to release cover
Press on edge of cover to snap closed
FOR PATIENT CARE AREAS
Power
Protection
1
2
Insert plug blade to release cover
Press on edge of cover to snap closed
5306M
5304M
(Top View | Vue du dessus
Vista superior)
PK-93031-10-02-0A
5304M-2N
5306M-2N
5304M-2S
5306M-2S
53A4M-2S
53A6M-2S
ENGLISH
Line |Ligne|Línea
125V AC/CA
Neutral|Neutre|Neutro
60HZ
Ground|Mise à la terre|Tierra
All Models | Tous les modèles| Todos los modelos
(Side View | Vue de profil| Vista lateral)
1.67 in
1.54 in
(4.24 cm)
(3.91 cm)
2.60 in
(6.60 cm)
13.80 in
(35.05 cm)
8.93 in
(22.68 cm)
MEDICAL GRADE
COMPLIANT WITH
NFPA 99 REQUIREMENTS
FOR PATIENT CARE AREAS
Power
10.80 in
Protection
(27.43 cm)
1
2
Insert plug blade to release cover
Press on edge of cover to snap closed
5302M
(Top View | Vue du dessus
Vista superior)

Publicité

loading

Sommaire des Matières pour Leviton 5302M-1N

  • Page 1 LIMITED LIFETIME WARRANTY AND EXCLUSIONS Leviton warrants to the original consumer purchaser and not for the benefit of anyone else that this product at the time of its sale by Leviton is free of defects in materials and workmanship under normal and proper use during the lifetime of the product.
  • Page 2 à la Manufacture Leviton du Canada Limitée, au soin du service de l'Assurance Qualité, 165 boul. Hymus, Pointe-Claire (Québec), H9R 1E9 (ou à Leviton Manufacturing Co., Inc., Att: Quality Assurance Department, 201 North Service Road,...

Ce manuel est également adapté pour:

5302m-1s53a2m-1s5304m-1n5304m-1s53a4m-1s5306m-1n ... Afficher tout