Intended Use; Approvals - Dräger X-act 5000 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

2.4

Intended use

The Dräger X-act 5000 is used for spot check measurements with
tubes, simultaneous Test-Sets, and sampling with sampling tubes and
systems. It is designed to draw the gas, which is to be examined or to
be sampled, through the respective tube or sampling system.
2.5

Approvals

The Dräger X-act 5000 is an air sampling pump for use in hazardous
locations. It is approved as Model GEP-0100. It is intrinsically safe
when used with one of the approved power supplies (see chap. 9 on
page 25).
Approval
CE marking
Approved according to 2014/30/EC and 2014/34/
EC
Europe/IECEx
I M1 / II 1G
Ex ia I Ma, Ex ia IIC T4 Ga
DEMKO 11 ATEX 1014422X
USA/Canada
UL/cUL:
Class I, Div. 1 Group A, B, C, D
Class II, Div. 1 Group F, G
+5 °C  Ta  +40 °C
4DC4
Dräger X-act 5000
NOTICE
i
i
The following items are not covered by the ATEX/IECEx/UL/
cUL approvals:
extension hose
adapter for extension hose
shoulder strap
resistance tube
adapter for sampling tubes
wall-mount charger
in-cab charger
The Dräger X-act 5000 complies with the FDA performance standards
for laser products, except for deviations pursuant to Laser Notice
No 50, dated June 24, 2007.
Description
11

Publicité

Table des Matières
loading

Table des Matières