Maintenance
4.11 Configuring the device settings
Briefly press
and
Monitoring.
To open the device settings:
1. Select
in measuring mode and confirm the
dialog.
2. Sign in with the required user level, if necessary.
3. Select and confirm Settings.
4.11.1 Activating day or night mode
1. Open the device settings.
2. Select and confirm Night mode / Day mode.
4.11.2 Changing the device language
1. Open the device settings.
2. Select Changing language....
3. Select and confirm the desired language.
4.11.3 Set date and time
1. Open the device settings.
2. Select Date & Time.
3. Select Set date format and then select and confirm
the date format.
4. Select Set date and then set and confirm the date.
5. Select Set time and then set and confirm the time.
The user must manually switch between summer
and winter time.
When using the X-dock maintenance station,
automatic time synchronisation is possible.
4.11.4 Activating silent mode
Silent mode can be activated for 15 minutes on the gas
detector. When silent mode is active, vibration and the
horn are deactivated. Permanent deactivation is
possible using the Dräger CC-Vision PC software.
Silent mode may only be permanently activated for
applications according to ATEX.
34
simultaneously to navigate to
5
Maintenance
5.1 Maintenance intervals
Test
Inspection and mainte-
nance by specialists.
For inspections and maintenance, see:
– EN 60079-29-2 – Guidelines for selection,
installation, use and maintenance of instruments for
the detection and measurement of flammable gases
and oxygen
– EN 45544-4 – Electrical apparatus used for the
direct detection and direct concentration
measurement of toxic gases and vapours – Part 4:
Guide for selection, installation, use and
maintenance
– National regulations
5.2 Calibration intervals
Observe the relevant specifications in the Sensor
Handbook or in the instructions for use/data sheets of
the DrägerSensors installed.
Recommended calibration intervals for DrägerSensors:
DrägerSensor
CatEx, O
, H
S, H
S LC,
2
2
2
CO LC, SO
, NO
2
2
5.3 Perform bump test
The bump test can be performed as follows:
– Bump test with the wizard (quick bump test)
– Bump test with X-dock (quick or extended bump
test)
Dräger recommends using the extended bump test
for cross calibrations (Dräger X-dock instructions for
use).
Instructions for use
Interval
Every 12 months
Calibration interval
Every 6 months
|
X-am® 3500