For Your Safety; Instructions For Use - Dräger DrugTest 5000 TTK Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 7
en
1

For your safety

1.1
Strictly follow the Instructions for Use
All usage of the Dräger DrugTest
understanding and strict observance of these Instructions for
Use.
The components of the Dräger DrugTest 5000 TTK only are
intended for the described use.
2
Description
Each packing unit contains:
Dräger DrugTest 5000 TTK, each packed in a sealed
foil pouch. The exact number is given on the label on the
outer pack.
Every foil pouch contains a desiccant pouch and a Dräger
DrugTest 5000 TTK comprising:
a. a test cassette with integrated sample collector
b. a safety cap with cartridge

Instructions for Use

2.1
Required materials that are not included
Dräger DrugTest 5000 Analyzer (order no. 83 19 900) for
processing and analysing test cartridges.
2.2
Additional recommended materials
Dräger Mobile Printer (order no. e.g. 83 19 310)
Disposable gloves (e.g. latex or nitrile gloves)
2.3
Intended use
The Dräger DrugTest 5000 TTK is used for training in the use
of the Dräger DrugTest 5000 test system that consists of the
Dräger DrugTest 5000 STK and the Dräger DrugTest 5000
Analyzer.
2.4
Explanation of symbols
Strictly follow the Instructions for Use!
Manufacturer
Batch
LOT
Temperature limitation
Expiry date
Contents sufficient for <n> tests
Measurement mode: fast (THC)
Measurement mode: sensitive (THC)
2.5
Mode of operation
The Dräger DrugTest 5000 TTK does not contain any
components suitable for use as drug evidence, but is externally
the same design as the Dräger DrugTest 5000 STK. This
means that the Dräger DrugTest 5000 TTK can be used to
demonstrate and test all usage and analysis stages on the test
system from oral fluid sampling through to the display of the
results on the Analyzer. Internally, the test cartridge bears a
1) Dräger DrugTest is a registered trademark of Dräger.
Dräger DrugTest 5000 TTK
®
1)
5000 TTK
requires full
special ID that is detected by the Analyzer. Depending on
whether a test cartridge with a red or yellow cap is used, the
Analyzer gives a positive or negative test result after completing
the regular test stages.
Using a test cartridge with a red cap (identification on
the label VERSION: TRAINING (+)) will deliver a positive
result.
Using a test cartridge with a yellow cap (identification on
the label VERSION: TRAINING (-)) will deliver a negative
result.
The yellow cartridges used with the training test cartridges,
do not contain any buffer solution. The risk of fluid escaping
is avoided, even following incorrect handling.
2.6
Precautions and warnings
NOTICE
i
i
For hygiene reasons, gloves must be worn during all
handling of the Dräger DrugTest 5000 Test System
and all samples. Do not touch the test cassette with
bare hands before and after sampling and follow the
current hygiene rules.
Follow handling and disposal procedures properly.
Open the foil pouch only immediately before use to prevent
contamination of the sample collector.
Do not use the Dräger DrugTest 5000 TTK if the foil pouch
is not intact.
Do not use the Dräger DrugTest 5000 TTK if the expiration
date on the pouch has passed. The expiration date is
formatted as YYYY-MM, e.g. 2011-01 means the Dräger
DrugTest 5000 TTK should not be used after the end
of January, 2011.
The Dräger DrugTest 5000 TTK can only be processed
using the Dräger DrugTest 5000 Analyzer!
Only use test cartridges with the corresponding yellow
cartridges. When using a blue cartridge, which is for
running actual drug tests, liquid may leak out.
3
Running the test simulation
The handling and analysis stages to be carried out using the
Dräger DrugTest 5000 TTK conform in full to the procedure for
the Dräger DrugTest 5000 test system, see the corresponding
Instructions for Use for the Dräger DrugTest 5000 STK and the
Dräger DrugTest 5000 Analyzer.
3.1
Test preparation and sampling
Switch on Analyzer.
3.1.1
Collecting the oral fluid sample
NOTICE
i
i
Observe the relevant Instructions for Use for the
Dräger DrugTest 5000 STK.
1. Remove the safety cap with cartridge from the sample
collector of the test cassette (Fig. 1) and give the test
cassette to the test subject. Keep the safety cap with
cartridge!
5

Publicité

Table des Matières
loading

Table des Matières