Philips M1191B Manuel D'utilisation page 8

Les capteurs de spo2 réutilisables
Masquer les pouces Voir aussi pour M1191B:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 12
Applying Glove Style Sensors
Adult Patients (>50kg)
Apply M1191B or M1191BL Sensor to any finger
except thumb.
Pediatric Patients (15-50kg)
Apply M1192A Sensor to any finger except thumb.
Infant Patients (4-15kg)
Apply M1195A Sensor to any finger (except thumb), or to any toe. Finger
or toe diameter should measure between 7-8mm (0.27-0.31").
Step
1
Select appropriate sensor for patient size (defined above).
2
Position sensor over patient's finger. If applying an adult finger
sensor, be sure the sensor cable is positioned ON TOP of the fin-
ger/hand as shown above.
3
Patient fingertip should touch but not protrude from end of sensor.
If necessary, trim the fingernail to correctly position the sensor.
4
If applying an adult finger sensor, secure the cable to BACK OF
HAND with M1627A wristband (supplied only with adult finger
sensors.)
5
Plug sensor into instrument (or into adapter cable if needed).
6
Inspect and change the sensor application site periodically.
Applying the Glove Style Sensor
6
adult application
shown

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

M1191blM1192aM1193aM1194aM1195a

Table des Matières