Philips M1191B Manuel D'utilisation page 10

Les capteurs de spo2 réutilisables
Masquer les pouces Voir aussi pour M1191B:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 12
Cleaning and Low-Level Disinfection
These reusable sensors should be cleaned and disinfected, but never
sterilized. Follow the procedure outlined below.
Warnings
Use only validated cleaning agents and disinfectants listed below; use
of others that are not validated may damage the sensor or its connect-
ing wires and shorten product lifetime or cause safety hazards.
Select disinfectants carefully as some have very similar names but
completely different compositions.
Do not immerse the sensor connector in any of the cleaning solutions,
disinfectant, or other liquid (the sensor and cable housing may be
immersed, but not the connector).
Do not soak sensors in disinfectants longer than specified by the dis-
infectant manufacturer.
Do not sterilize or autoclave the sensors.
Validated Cleaning Agents
- Mild Detergent
Validated Disinfectants
®
Metricide
28
®
Metricide
Plus 30
®
Terralin
Liquid
®
Incidin
Liquid
Step
1
Clean sensor following instructions supplied with cleaning agent.
2
Disinfect sensor following instructions supplied with disinfectant.
3
Rinse sensor and cable (but not the connector) in water, wiping it dry with a
clean cloth, leaving to dry completely. If you notice any deterioration or
damage to sensor or cable, dispose the sensor immediately.
- Salt Solution (1%)
®
Cidex
Formula 7
®
Cidex
OPA
®
Cidex
Plus
®
Omnicide
Cleaning and Low-Level Disinfection
Kohrsolin
Mucapur
Isopropanol (70%) or
Isopropanol Wipe (70%)
28
8
®
(2%)
®
-CD (1%)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

M1191blM1192aM1193aM1194aM1195a

Table des Matières