5.2
Inserting or removing the gas detector
into or from a module
To insert the instrument in the module:
1. If necessary, push the lock up slightly and open the module
cover upwards.
2. Place the instrument in the corresponding module. The
temperature of the gas detector must be >0 °C.
3. Close the module cover and make sure it engages on both
sides.
The instrument will be detected automatically.
X-am 125+/8000+ module with charging function only:
After the instrument is inserted, the charging status is
displayed for approx. 5 seconds via the charging status
LED.
The charging function starts automatically approx.
15 minutes after the last test.
To remove the instrument from the module:
1. Push the lock up slightly and open the module cover
upwards.
2. Remove the instrument.
5.3
Station self-test
A self-test is carried out:
When the station is started up.
When the last successful self-test was more than 24 hours
ago and a test is being carried out.
The station is tested for leaks, pump function, and the software
version of the individual modules and the master.
Dräger X-dock 5300/6300/6600
5.4
i
i
i
i
2
1
00233286.eps
i
i
The following tests are preconfigured:
Test 1: QT
Test 2: EXT
Test 3: CAL
1. If necessary, open the test gas cylinders.
2. If necessary, switch on the X-dock.
3. Conduct a visual inspection of the instruments (see
chapter 5.1 on page 26).
4. Insert the instruments into the modules (see chapter 5.2 on
page 27).
If Single mode is activated:
The default test is started automatically when the module
flap closes.
The Status LED flashes blue.
The individual test phases are displayed.
If Favorites mode is activated:
1. Select the desired test from favourites bar.
2. If the test requires a higher authorisation level, log in at the
station with the respective authorisation level (see
chapter 4.5 on page 23).
Test will be started automatically.
The status LED flashes blue.
The individual test phases are displayed.
Conducting a test
NOTICE
When using a gas >100 %LEL for the bump test,
Dräger recommends opening the respective test gas
cylinder only for the duration of the test or calibration,
and selecting the "Flush" test option (see Technical
manual).
NOTICE
Single mode is activated by default.
Several tests can be started and executed in parallel in
the Single mode.
A failure of a LED, horn or vibration test results in a
negative evaluation of the overall test, and thus in the
locking of the respective gas measurement system.
A testing of the sensor reserve is only performed with
sensors that support this function. The results are
displayed under test details and give information about
the state of the sensor.
NOTICE
If an X-am 8000 remains off for longer than 21 days
and is not charged, deep sleep mode is activated
automatically. In deep sleep mode, the gas detector
can no longer be switched on using the Dräger
CC-Vision PC software or the Dräger X-dock. In this
case, the gas detector must be switched on manually.
Quick bump test including checking the alarm
elements.
Extended bump test incl. zero-point check and
checking the alarm elements.
Calibration and checking the alarm elements.
Use
27