Dräger X-dock 5300 Notice D'utilisation page 25

Masquer les pouces Voir aussi pour X-dock 5300:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 35
4.6.1
Settings
1. Select
> Station gas configuration > Settings.
Select test gas for X-am 2000
Fresh air inlet setting
Expert settings
Gas cylinder level indicator
For the X-am 2000, it is possible to choose between three
different test gases that are also used for calibration and
testing. The three options are methane (CH
and pentane (C
H
). The sensor is calibrated with different
5
12
sensitivity levels depending on the gas selected. More
information on this can be found in the relevant sensor data
sheets.
NOTICE
i
i
The corresponding gas must be connected to one of
the gas inlets and set in the gas configuration.
There is also the option of setting an "increased sensitivity"
option for propane and pentane. This option artificially
increases the sensitivity to calibrate the sensors so that these
have approximately a nonane sensitivity level (in other words,
a sensitivity level as if they were calibrated for nonane). More
information on the subject of cross-sensitivity calibration can
be found in the relevant sensor data sheets.
To select the test gas for X-am 2000:
1. Select X-am 2000.
2. Select the required test gas from the list.
The following selection is available:
Methane - CH
(default setting)
4
Propane - C
H
3
8
Pentane - PENT
For propane and pentane, the "Increased sensitivity"
(vapour sensitivity) option can also be activated.
3. Confirm selection with OK.
Dräger X-dock 5300/6300/6600
), propane (C
H
)
4
3
8
To set the fresh air input:
1. Select Fresh air.
2. Select desired setting:
Supplied by pump - fresh air inlet (default setting)
Supplied by compressed air input - compressed air
inlet
3. Confirm selection with OK.
The following settings can be made in the expert menu:
Ignore max. concentration for quick bump test
Set test behaviour for missing test gases
To ignore the concentration limit specified by Dräger for the
quick bump test:
1. Select Ignore max. conc. for BTQ.
2. Activate checkbox (default setting: deactivated).
3. Confirm selection with OK.
Activating this feature allows users to use higher test gas
concentration for the quick bump test than recommended by
Dräger.
WARNING
!
This feature must only be activated by trained and
qualified personnel since wrongly selected test gas
concentrations may lead to a positive test result
although the gas detector's warning comes too late.
To set the test behaviour with missing test gases:
1. Select Cancel the test on missing gas.
2. Activate checkbox (default setting: activated).
3. Confirm selection with OK.
This function can be used to set whether or not a test or
calibration is carried out when a required test gas is not
connected.
WARNING
!
If this function is deactivated, the corresponding
channel is not tested or calibrated.
Basics
25

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

X-dock 5300 x-amX-dock 5300 pacX-dock 5300 x-am 8000X-dock 6300X-dock 6600

Table des Matières