Dräger DrugTest 5000 Manuel Technique page 23

Masquer les pouces Voir aussi pour DrugTest 5000:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Error
Description
Code
22001
(QA
QA Check failed
Check
failed)
Test cassette cannot be
20006
processed
Login as operator failed
22020
(wrong password was
entered)
22040
New operator: operator list full Delete one operator
New operator: operator
22041
already exists
New operator: password
22043
already used or 2nd pass-
word is 1st password
Change operator password:
22050
wrong old password
Change operator password:
22052
new password already used
Change operator password:
22053
2nd new password is 1st new
password
30000
Service: Service expired
Service: Service to be carried
30001
out soon
User error: Measurement
30013
cancelled by operator
User error: Temperature
30022
inside of the instrument is out
of working range
Dräger DrugTest 5000 Test System
Corrective Action
Device cannot be used. Maintenance is
required.
Ensure correct QA cassette is used
Enter correct password
Use another name
Choose a different password or enter
correct new password
Enter correct password
Choose different password from existing
one
Enter correct new password
Device can be used, but maintenance
should be carried out soon.
Device can be used, but maintenance
should be carried out soon.
No error
Usage within temperature range
according to IFU
Text on Screen
QA-Test failed! Remove cassette.
Close door.
Error when reading QA-cassette!
Remove cassette.
Invalid password. Login cancelled.
Too many operators.
No further entry possible.
Transfer new operator list.
The operator has already been set up
and could not be stored again.
Invalid password.
Please choose another password.
Wrong password entered.
Please try again.
Invalid password.
Please choose another password.
Two different passwords have been
entered.
Enter password again.
New operator not assigned.
"Service 1" expired!
Please carry out maintenance.
"Service 1" to be carried out soon.
Do you really want to cancel the current
measurement?
Cancel measurement by pressing OK or
return to measurement.
Device temperature is beyond operational
range.
Wait for equalization of temperature or
switch off device.
Troubleshooting
23