Troubleshooting
Error
Description
Code
10811
(during
Low Battery warning
measur
ement)
10811
(before
Low Battery warning
measur
ement)
User error: Door is still open
20000
and should be closed
User error: Cartridge and/or
20001
cassette not removed
User error: Entered date is not
20002
plausible
User error: Entered time is not
20003
plausible
User error: Door was opened
20004
while measurement was
running
User error: Shelf life of test
20005
cassette has expired
Database failure: Test
20006
cassette cannot be processed
User error: Wrong QA
20011
cassette was used
User error: QA test was
20012
cancelled
User error: Printer paper is
21200
missing or printer flap open
User error: Communication
21201
error (no printer found)
22000
Service interval expired
22000
(QA
QA Check failed
Check
failed)
22001
QA Check expired
22
Corrective Action
Charge battery
Charge battery
Close door
Remove cassette and close door
Enter correct date
Enter correct time
Keep door closed during measurement
Use valid test cassette
Update software
Ensure correct QA cassette is used
No error
Check paper and printer cap
Switch on printer, re-check positioning of
printer. If error persists, service required
Device cannot be used. Maintenance is
required
Device cannot be used. Maintenance is
required.
Run QA-Test
Text on Screen
Rechargeable battery is completely
depleted.
Connect mains adaptor or finish
measurement before device will be
switched off.
Rechargeable battery is completely
depleted.
Connect mains adaptor or switch off
device.
The Analyzer's sliding door is still open.
Close door.
There is still a cassette or cartridge
in the device.
Remove cassette and cartridge and close
the door.
Entered date is not valid.
Enter valid date.
Date not set.
Entered time is not valid.
Enter valid time.
Time not set.
Door has been opened during
measurement.
Measurement has been cancelled.
Remove cassette and cartridge and close
the door.
Shelf life of the cassette has expired.
Measurement has been cancelled.
Remove cassette and cartridge and close
the door.
Panel-Type cannot be measured with this
device.
Measurement has been cancelled.
Remove cassette and cartridge and close
the door.
Wrong QA-cassette inserted!
Remove cassette.
QA-Test cancelled. Counter has not been
reset. Remove cassette. Close door
Printer out of paper or printer cover not
closed. Insert paper, close cover and press
OK.
Infrared connection between printer
and Analyzer failed.
Switch on printer and align IR port with
Analyzer (place printer on the left).
"Service 1" expired!
Measurement not possible!
Please carry out maintenance.
"QA Check" QA-Test failed!
Measurement not possible!
Please carry out maintenance.
"QA-Check" expired!
Measurement not possible!
Run QA-Test(s)!